Functional check
3-15
Document no. 2044677-001
−
Waveform 2 - P1
−
Waveform 3 - P2
Invasive Pressures
- P1 ‘ART’ Setup
- Label - ART
- P2 ‘CVP’ Setup - Label - CVP
Others
-SPO2 Setup - Pleth Scale - AUTO
3.2.2 Display
1.
Check that the picture on the screen is displayed properly.
"
3.2.3 Keyboard(s)
Tests with the Command Board:
−
Press the
Monitor Setup
key. Turn the ComWheel in both directions and check that the
cursor in the menu moves correspondingly. Select
Normal Screen
and check that the
menu disappears from the screen.
Check the rest of the menu keys by pressing them one by one.
"
3.2.4 Frame unit
1.
Check that the clock on the screen shows correct time. Readjust the time and date, if
necessary.
"
3.2.5 Extension Module with CO
2
measurement
Wait until the message ‘Calibrating gas sensor’ disappears from the screen.
1.
Block the tip of the sampling line with your finger and check that the message ‘Sample
line blocked’ appears on the monitor screen within 30 seconds.
2.
Detach the Mini D-fend and check that the message ‘Check D-fend’ appears on the
monitor screen within 30 seconds.
Breathe to the sampling line briefly. Check that the CO2 information is updated on the
screen.
"
3.2.6 Multiparameter Hemodynamic Modules
ECG and RESP measurements
1.
Connect an ECG cable to the module. Connect the cable leads to a patient simulator.
Check that all ECG and impedance respiration information is shown on the monitor
screen as configured on the simulator.
Turn the simulator off. Check that the ëAsystoleí and ëApneaí messages are displayed.
Summary of Contents for b30
Page 2: ......
Page 12: ...B30 Patient Monitor 8 Document no 2044677 001...
Page 13: ...1 Introduction...
Page 14: ......
Page 61: ...2 Installation...
Page 69: ...3 Maintenance...
Page 70: ......
Page 105: ...4 Troubleshooting...
Page 106: ......
Page 125: ...5 Service Menu...
Page 126: ......
Page 158: ...B30 Patient Monitor 5 32 Document no 2044677 001...
Page 159: ...6 Field replaceable unit...
Page 160: ......
Page 166: ...B30 Patient Monitor 6 6 Document no 2044677 001 1 6 Frame for B30 Frame for B30...
Page 183: ...Frame disassembly 6 23 Document no 2044677 001...
Page 190: ...B30 Patient Monitor 6 30 Document no 2044677 001...
Page 191: ...7 Technical specification...
Page 192: ......
Page 198: ...B30 Patient Monitor 7 6 Document no 2044677 001...
Page 199: ...8 N F C REC Module...
Page 200: ......
Page 238: ...B30 Patient Monitor A 6 Document no 2044677 001...
Page 251: ......