![GE b30 Technical Reference Manual Download Page 117](http://html1.mh-extra.com/html/ge/b30/b30_technical-reference-manual_4013170117.webp)
E-PSM(P)W Troubleshooting
4-11
Document no. 2044677-001
Cuff ID not working
1. Defective cuff ID holes in the
NIBP cuff hose
- Replace NIBP cuff hose.
2. NIBP flex board connector
wrongly connected
- Check that the NIBP flex board
connector is properly connected to
the STP input board: all pins have to
be connected.
3. Cuff ID switches defective
- To check the switches, attach a NIBP
cuff hose without the cuff ID and
check that the message ‘Select
inflation limit’ appears. If not, replace
the Front Panel Unit.
‘Air leakage’ message
1. Hose or cuff leaking. Reasons:
1. Replace cuff
- cuff damaged
- Replace cuff.
- cuff connector damaged
- Replace cuff connector (if the fault is
in hose connector).
- O-ring damaged or missing
- Replace O-ring.
- hose double connector damaged - Replace NIBP cuff hose.
2. Hose and cuff OK. Reasons:
2. Connect or replace tube
- leakage in the tubes connecting
the patient connector panel and
the module
- Check the tubes.
- leakage inside the module
- Replace the whole tubing.
- tube disconnected or damaged
- Fix connections.
- manifold leaking
- Replace the manifold.
- tubes or valve(s) damaged
- Replace tubes/valve(s).
‘Unable to measure Sys’
message
Systolic blood pressure probably
higher than the inflation pressure or
artifacts.
Automatic retrial with increased
pressure.
Problem
Cause
What to do
Summary of Contents for b30
Page 2: ......
Page 12: ...B30 Patient Monitor 8 Document no 2044677 001...
Page 13: ...1 Introduction...
Page 14: ......
Page 61: ...2 Installation...
Page 69: ...3 Maintenance...
Page 70: ......
Page 105: ...4 Troubleshooting...
Page 106: ......
Page 125: ...5 Service Menu...
Page 126: ......
Page 158: ...B30 Patient Monitor 5 32 Document no 2044677 001...
Page 159: ...6 Field replaceable unit...
Page 160: ......
Page 166: ...B30 Patient Monitor 6 6 Document no 2044677 001 1 6 Frame for B30 Frame for B30...
Page 183: ...Frame disassembly 6 23 Document no 2044677 001...
Page 190: ...B30 Patient Monitor 6 30 Document no 2044677 001...
Page 191: ...7 Technical specification...
Page 192: ......
Page 198: ...B30 Patient Monitor 7 6 Document no 2044677 001...
Page 199: ...8 N F C REC Module...
Page 200: ......
Page 238: ...B30 Patient Monitor A 6 Document no 2044677 001...
Page 251: ......