B30 Patient Monitor
3-4
Document no
.
2044677-001
2
Electrical Safety Tests
2.1 General
Electrical safety tests provide a method of determining if potential electrical health hazards to
the patient or operator of the device exist.
2.2 Recommendations
GE recommends that you perform all safety tests presented in this chapter.
•
Upon receipt of the device (monitor and its associated equipment).
•
Every twelve months thereafter planned maintenance
•
Each time the main enclosure is disassembled or a circuit board is removed, tested,
repaired, or replaced
These instructions are intended for every component in the system. GE recommends that the
qualified personnel performing the tests.
2.3 Test Equipment
The recommended test equipment required to perform electrical safety tests is listed below.
Perform electrical safety tests using an electrical safety analyzer per IEC 60601-1, UL 60601-1,
EN 60601-1 or CSA C22.2 No. 601.1. The schematics in the section provide a general
understanding of the test equipment. Actual configuration of test equipment may vary.
The patient monitor being tested should be placed on an insulating surface.
2.4 Power Outlet Test
Verify that the power outlet is wired correctly per the countryís electrical code standard before
starting the following electrical safety tests. The results of the following tests will be inaccurate
unless a properly wired power outlet is used. Use only non-isolated power outlets when
performing safety tests.
2.5 Power cord and plug
Verify the power cord being used with the patient monitor is good. The following are a couple
of things to check for in this regard:
Item
Specification
Leakage Current Tester
Equivalent to the circuits
shown
Digital Multimeter (DMM) (optional based on
leakage tester used and locality)
AC volts, ohms
Ground Bond Tester
0 – 1 ohm
ECG Test Body
All leads together
GE and Nellcor SPO2 Test Body
2006646-001
Summary of Contents for b30
Page 2: ......
Page 12: ...B30 Patient Monitor 8 Document no 2044677 001...
Page 13: ...1 Introduction...
Page 14: ......
Page 61: ...2 Installation...
Page 69: ...3 Maintenance...
Page 70: ......
Page 105: ...4 Troubleshooting...
Page 106: ......
Page 125: ...5 Service Menu...
Page 126: ......
Page 158: ...B30 Patient Monitor 5 32 Document no 2044677 001...
Page 159: ...6 Field replaceable unit...
Page 160: ......
Page 166: ...B30 Patient Monitor 6 6 Document no 2044677 001 1 6 Frame for B30 Frame for B30...
Page 183: ...Frame disassembly 6 23 Document no 2044677 001...
Page 190: ...B30 Patient Monitor 6 30 Document no 2044677 001...
Page 191: ...7 Technical specification...
Page 192: ......
Page 198: ...B30 Patient Monitor 7 6 Document no 2044677 001...
Page 199: ...8 N F C REC Module...
Page 200: ......
Page 238: ...B30 Patient Monitor A 6 Document no 2044677 001...
Page 251: ......