B30 Patient Monitor
8-18
Document no
.
2044677-001
5
N-FCREC Module disassembly and
reassembly
CAUTION
When reassembling the module, make sure that the tubes and cables are not
pinched between the boards and the cover.
5.0.1 Instructions after replacing MiniC unit
After replacing the MiniC unit:
•
perform the sampling system leak test
•
perform the occlusion test
•
perform the gas calibration
5.0.2 Before disassembly
NOTE: Wear a grounded, antistatic wristband when handling PC boards. Electrostatic
discharge may damage components on the board.
Disconnect the Extension Module from the monitor.
Tools needed
- screwdrivers, TORX; T8, T10, T15
- crosshead screwdriver; M15
- flat blade screwdriver
- pinchers
- antistatic wristband
Disassemble the Extension Module
1. Remove the Mini D-fend.
2. Open the recorder unit paper cover and remove
the paper roll, if installed.
Summary of Contents for b30
Page 2: ......
Page 12: ...B30 Patient Monitor 8 Document no 2044677 001...
Page 13: ...1 Introduction...
Page 14: ......
Page 61: ...2 Installation...
Page 69: ...3 Maintenance...
Page 70: ......
Page 105: ...4 Troubleshooting...
Page 106: ......
Page 125: ...5 Service Menu...
Page 126: ......
Page 158: ...B30 Patient Monitor 5 32 Document no 2044677 001...
Page 159: ...6 Field replaceable unit...
Page 160: ......
Page 166: ...B30 Patient Monitor 6 6 Document no 2044677 001 1 6 Frame for B30 Frame for B30...
Page 183: ...Frame disassembly 6 23 Document no 2044677 001...
Page 190: ...B30 Patient Monitor 6 30 Document no 2044677 001...
Page 191: ...7 Technical specification...
Page 192: ......
Page 198: ...B30 Patient Monitor 7 6 Document no 2044677 001...
Page 199: ...8 N F C REC Module...
Page 200: ......
Page 238: ...B30 Patient Monitor A 6 Document no 2044677 001...
Page 251: ......