![GE b30 Technical Reference Manual Download Page 82](http://html1.mh-extra.com/html/ge/b30/b30_technical-reference-manual_4013170082.webp)
B30 Patient Monitor
3-12
Document no
.
2044677-001
2.10 Patient (sink) leakage current test (mains voltage on
the applied part)
This procedure only applies to Class I (grounded/earthed) equipment, and measures the
leakage current from a mains voltage source into the ECG/RESP connector or the SpO
2
connector.
The patient connector test body shorts all signals in the connector together. Refer to the
instructions contained with the safety analyzer to perform this test. Connect the ECG/RESP Test
Body to the green connector of the device under test.
Refer to the instructions contained with the safety analyzer to perform each test.
NOTE: The MD represents the network and voltage measuring instrument and its frequency
characteristics per IEC 60601-1.
NOTE: Per IEC 60601-1, the impedance to protect the circuitry and the person performing the
test, but low enough to accept currents higher than the allowable values of the LEAKAGE
CURRENT to be measured.
(1) 1. Configure leakage tester as follows:
•
Polarity - NORMAL
•
Neutral - OPEN
•
GND (Earth) - CLOSED
WARNING
Shock hazard. The following step causes high voltage at the test body. Do
not touch the test body.
(2) Power on device under test.
(3) Read and record leakage current indicated on the tester.
(4) Change leakage tester switches to:
•
Polarity - NORMAL
•
Neutral - OPEN
•
GND (Earth) - CLOSED
(5) Read and record the current leakage indicated on the tester.
Power Cord
Leakage Tester
NORM
RVS
ECG Test Body
or ECG Cable
or SPO2 Test Body
HIGH
LOW
GND
1K
0.15µF
10
Power Cord
Device
Under
Test
DMM set to measure AC voltage
GND
DMM
Closed
(Keep cable length as
short as possible.)
120K
Summary of Contents for b30
Page 2: ......
Page 12: ...B30 Patient Monitor 8 Document no 2044677 001...
Page 13: ...1 Introduction...
Page 14: ......
Page 61: ...2 Installation...
Page 69: ...3 Maintenance...
Page 70: ......
Page 105: ...4 Troubleshooting...
Page 106: ......
Page 125: ...5 Service Menu...
Page 126: ......
Page 158: ...B30 Patient Monitor 5 32 Document no 2044677 001...
Page 159: ...6 Field replaceable unit...
Page 160: ......
Page 166: ...B30 Patient Monitor 6 6 Document no 2044677 001 1 6 Frame for B30 Frame for B30...
Page 183: ...Frame disassembly 6 23 Document no 2044677 001...
Page 190: ...B30 Patient Monitor 6 30 Document no 2044677 001...
Page 191: ...7 Technical specification...
Page 192: ......
Page 198: ...B30 Patient Monitor 7 6 Document no 2044677 001...
Page 199: ...8 N F C REC Module...
Page 200: ......
Page 238: ...B30 Patient Monitor A 6 Document no 2044677 001...
Page 251: ......