Service procedures
8-15
Document no. 2044677-001
Sampling flow (ml/min)
135...165
If necessary, readjust the sampling flow:
Select
Sample gain adj
from the menu.
To increase the sampling flow, turn the ComWheel counterclockwise. To decrease the
flow, turn the ComWheel clockwise.
A change of 0.050 in the Gain value changes the flow approximately 7.5 ml/min.
After you have changed the gain, wait until the Sample Flow value on the screen returns
near to the original, then check the flow meter reading again.
"
12. Working pressure
Check that the Amb-Work value in the service menu is within the following range:
Amb-Work (mmHg)
20...50
"
13. Gas calibration
Airway Gas
- Gas Calibration
NOTE: The calibration should not be performed before 30 minutes warm-up time.
Use calibration gas 755580 (5% CO
2
, about 20% O
2
) for calibrating the module with CO
2
measurement.
"
14. Occlusion detection
Block the tip of the sampling line with your finger and check that the message ‘Sample
line blocked’ appears on the monitor screen within 60 seconds.
"
15. Check D-fend
Detach the mini D-fend and check that the message ‘Check D-fend’ appears on the
monitor screen within 30 seconds.
"
16. Apnea detection
Reattach the mini D-fend. Simulate at least 5 breaths by feeding calibration gas into the
sampling line. Check that the shown gas information is correct.
Check that the monitor shows the message ‘Apnea’ within 30 seconds after you have
stopped feeding the gas.
"
Summary of Contents for b30
Page 2: ......
Page 12: ...B30 Patient Monitor 8 Document no 2044677 001...
Page 13: ...1 Introduction...
Page 14: ......
Page 61: ...2 Installation...
Page 69: ...3 Maintenance...
Page 70: ......
Page 105: ...4 Troubleshooting...
Page 106: ......
Page 125: ...5 Service Menu...
Page 126: ......
Page 158: ...B30 Patient Monitor 5 32 Document no 2044677 001...
Page 159: ...6 Field replaceable unit...
Page 160: ......
Page 166: ...B30 Patient Monitor 6 6 Document no 2044677 001 1 6 Frame for B30 Frame for B30...
Page 183: ...Frame disassembly 6 23 Document no 2044677 001...
Page 190: ...B30 Patient Monitor 6 30 Document no 2044677 001...
Page 191: ...7 Technical specification...
Page 192: ......
Page 198: ...B30 Patient Monitor 7 6 Document no 2044677 001...
Page 199: ...8 N F C REC Module...
Page 200: ......
Page 238: ...B30 Patient Monitor A 6 Document no 2044677 001...
Page 251: ......