E-PSM(P)W Troubleshooting
4-15
Document no. 2044677-001
3.4 Impedance respiration
Problem
Cause
What to do
No resp trace
Waveform not selected on the screen
Press the Monitor Setup key and make
adjustments.
Module not plugged in correctly
Re-plug the module.
Unacceptable resp
waveform
Poor electrode or poor electrode skin
contact
Electrodes from different
manufacturers are used. Too
much/little gel is used.
Poor electrode condition
Electrodes are dried out.
Improper site of electrodes
Check that electrodes are not placed
over bones, active muscles, or layers of
fat.
Improper skin preparation
Remove body hair. Clean attachment
site carefully with alcohol.
Faulty/ dirty ECG cable.
Change new cable.
Message: ‘SMALL RESP
CURVE’
Respiration signal is very small
With 3-lead cable try another lead
connection I, II, III or try 5-lead cable.
Message: ‘APNEA ALARM’,
and respiration waveform
normal
Respiration source is CO
2
Check respiration source and change it
to correct one.
Summary of Contents for b30
Page 2: ......
Page 12: ...B30 Patient Monitor 8 Document no 2044677 001...
Page 13: ...1 Introduction...
Page 14: ......
Page 61: ...2 Installation...
Page 69: ...3 Maintenance...
Page 70: ......
Page 105: ...4 Troubleshooting...
Page 106: ......
Page 125: ...5 Service Menu...
Page 126: ......
Page 158: ...B30 Patient Monitor 5 32 Document no 2044677 001...
Page 159: ...6 Field replaceable unit...
Page 160: ......
Page 166: ...B30 Patient Monitor 6 6 Document no 2044677 001 1 6 Frame for B30 Frame for B30...
Page 183: ...Frame disassembly 6 23 Document no 2044677 001...
Page 190: ...B30 Patient Monitor 6 30 Document no 2044677 001...
Page 191: ...7 Technical specification...
Page 192: ......
Page 198: ...B30 Patient Monitor 7 6 Document no 2044677 001...
Page 199: ...8 N F C REC Module...
Page 200: ......
Page 238: ...B30 Patient Monitor A 6 Document no 2044677 001...
Page 251: ......