![GE b30 Technical Reference Manual Download Page 236](http://html1.mh-extra.com/html/ge/b30/b30_technical-reference-manual_4013170236.webp)
B30 Patient Monitor
A-4
Document no
.
2044677-001
5.
Click
Open License
button, choose the license file you have. The
CPU Serial Number
and
Software Version
will be automatically imported through the license file.
6.
Press
Emergence
button to start software download.
NOTE: When the emergence download for the main software is done successfully,
waiting for 3 to 5 minutes until the B30 monitor restart automatically.
Verification and configuration
1.
Perform Factory Reset.
Monitor Setup
Install/Service
(16-4-34)
Service
(26-23-8)
Set/Test
Factory Reset
After the Factory Reset is completed, the B30 monitor will restart automatically.
2.
Set the time and date.
Monitor Setup
Time and Date
3.
If necessary, set the monitor’s network communication according to the used network
software.
Monitor Setup
Install/Service
(16-4-34)
Service
(26-23-8)
Frame
Network
Dri Config
DRI level
Central station software S-CNET01 -> DRI Level = 2001
Central station software S-CNET02 -> DRI Level = 2002
Central station software L-NET03 or L-NET05 -> DRI Level = 2003
NOTE: If the DRI level is changed, the B30 monitor will restart automatically. The
change will take effect only after restart.
4.
Verify that the software has been installed correctly.
Monitor Setup
Install/Service
(16-4-34)
Service
(26-23-8)
SW Version / Unit Id
5.
Verify that the active language is installed and activated correctly
Monitor Setup
Install/Service
(16-4-34)
Service
(26-23-8)
SW Settings
Country Settings
Summary of Contents for b30
Page 2: ......
Page 12: ...B30 Patient Monitor 8 Document no 2044677 001...
Page 13: ...1 Introduction...
Page 14: ......
Page 61: ...2 Installation...
Page 69: ...3 Maintenance...
Page 70: ......
Page 105: ...4 Troubleshooting...
Page 106: ......
Page 125: ...5 Service Menu...
Page 126: ......
Page 158: ...B30 Patient Monitor 5 32 Document no 2044677 001...
Page 159: ...6 Field replaceable unit...
Page 160: ......
Page 166: ...B30 Patient Monitor 6 6 Document no 2044677 001 1 6 Frame for B30 Frame for B30...
Page 183: ...Frame disassembly 6 23 Document no 2044677 001...
Page 190: ...B30 Patient Monitor 6 30 Document no 2044677 001...
Page 191: ...7 Technical specification...
Page 192: ......
Page 198: ...B30 Patient Monitor 7 6 Document no 2044677 001...
Page 199: ...8 N F C REC Module...
Page 200: ......
Page 238: ...B30 Patient Monitor A 6 Document no 2044677 001...
Page 251: ......