![background image](http://html1.mh-extra.com/html/ge/b30/b30_technical-reference-manual_4013170140.webp)
B30 Patient Monitor
5-14
Document no
.
2044677-001
4.1.2 Gases
Noise Meas
activates the noise
measurement.
Sample gain adj
adjusts the sampling
pump gain, i.e. for
adjusting the sample flow
measurement.
Pump ctrl
A manual control for the
sampling pump.
Zero valve ctrl
A manual control for the
zero valve.
Service Data
CO
2
% field
shows
real-time
concentrations.
CO
2
noise-%
is standard deviation of
concentration.
CO
2
/channels A-E
mV
field: signal is scaled
to mV.
Gain
: User gain. It is
scaled as (User
gain)/(Factory gain).
O2, N2O, AA
Not in use
ID
Not in use
ID unrel.
Not in use
MAC
Not in use
Sample Flow
is calculated from differential pressure and adjusted by the module.
Zero
value as
measured during initialization when the pump is off.
Gain
: sample flow measurement can
be calibrated by adjusting the gain.
Ambient
Ambient pressure is measured continuously.
Amb-Work
: ambient pressure - sampling
system internal pressure.
Fall time
Not in use
CO
2
-O
2
Delay
Not in use
Pump
Can be toggled ON/OFF. PWM output 0-100% is shown. Pump voltage is also shown.
Lamp
The state, PWM control, and current of the lamp are shown.
Fan
Not in use
Zero Valve
Can be toggled between the measurement state (MEAS) and the zeroing state (ZERO).
Occl Valve
Not in use
Temp
Temperatures measured by the module from TPX, CPU, and OM.
Time after power on
In minutes after power on.
Summary of Contents for b30
Page 2: ......
Page 12: ...B30 Patient Monitor 8 Document no 2044677 001...
Page 13: ...1 Introduction...
Page 14: ......
Page 61: ...2 Installation...
Page 69: ...3 Maintenance...
Page 70: ......
Page 105: ...4 Troubleshooting...
Page 106: ......
Page 125: ...5 Service Menu...
Page 126: ......
Page 158: ...B30 Patient Monitor 5 32 Document no 2044677 001...
Page 159: ...6 Field replaceable unit...
Page 160: ......
Page 166: ...B30 Patient Monitor 6 6 Document no 2044677 001 1 6 Frame for B30 Frame for B30...
Page 183: ...Frame disassembly 6 23 Document no 2044677 001...
Page 190: ...B30 Patient Monitor 6 30 Document no 2044677 001...
Page 191: ...7 Technical specification...
Page 192: ......
Page 198: ...B30 Patient Monitor 7 6 Document no 2044677 001...
Page 199: ...8 N F C REC Module...
Page 200: ......
Page 238: ...B30 Patient Monitor A 6 Document no 2044677 001...
Page 251: ......