![GE b30 Technical Reference Manual Download Page 46](http://html1.mh-extra.com/html/ge/b30/b30_technical-reference-manual_4013170046.webp)
B30 Patient Monitor
1-32
Document no
.
2044677-001
Software control
The software controls valves and a pump. In addition to the individual on/off signals for each
component there is a common power switch for the valves and the pump that can be used at
pump/valve failures.
In addition to external RS485 reset line, the microprocessor system is equipped with its own
power-up reset. See the section in the ECG board’s description: “RS485 communication”
Safety circuit
The NIBP board is equipped with an independent safety circuit to disconnect supply voltages
from the pump and the valves if the cuff has been pressurized longer than the preset
maximum measurement time, or if the pressure of the cuff is inflated over the specified
pressure limit. The maximum measurement time values and pressure limits for different
measurement modes have been specified in the technical specification section of this manual.
Pneumatics
The module has the following pneumatics parts:
1.
Intake air filter
; for preventing dust and other parts from entering the air pump and the
valves.
2.
Air pump
; for pumping the measuring pressure of the cuff.
3.
(Pulse) Valve
; for producing a linear pressure fall (bleeding) in order to measure the blood
pressure of the patient.
Note that in the service menu also names
Valve
and
Set valve
have been used for this
valve.
4.
Safety valve
; The safety valve is intended to be used for deflating the cuff in single fault
case, i.e. to prevent too long a measurement time or too high an inflation pressure of the
cuff.
Note that lso
Exh2 valve
has been used to designate the
Safety valve
in service menu.
5.
Main pressure sensor
; for measuring the pressure of the blood pressure cuff and the
pressure fluctuations caused by arterial wall movement.
6.
Safety pressure sensor;
for detecting the , cuff loose, cuff occlusion situations, etc. and
for recognizing the pressure sensor fault.
7.
Cuff connector;
for connection and hose identification.
Summary of Contents for b30
Page 2: ......
Page 12: ...B30 Patient Monitor 8 Document no 2044677 001...
Page 13: ...1 Introduction...
Page 14: ......
Page 61: ...2 Installation...
Page 69: ...3 Maintenance...
Page 70: ......
Page 105: ...4 Troubleshooting...
Page 106: ......
Page 125: ...5 Service Menu...
Page 126: ......
Page 158: ...B30 Patient Monitor 5 32 Document no 2044677 001...
Page 159: ...6 Field replaceable unit...
Page 160: ......
Page 166: ...B30 Patient Monitor 6 6 Document no 2044677 001 1 6 Frame for B30 Frame for B30...
Page 183: ...Frame disassembly 6 23 Document no 2044677 001...
Page 190: ...B30 Patient Monitor 6 30 Document no 2044677 001...
Page 191: ...7 Technical specification...
Page 192: ......
Page 198: ...B30 Patient Monitor 7 6 Document no 2044677 001...
Page 199: ...8 N F C REC Module...
Page 200: ......
Page 238: ...B30 Patient Monitor A 6 Document no 2044677 001...
Page 251: ......