Temp error
shows the status of the temperature test. No errors found show the
status (OFF) and errors found (ON).
Protect key
and
Protect mode
show normally ON but can be turned to OFF for
the temperature calibration in Calibration Menu.
Configuration
shows the chosen module configuration: TP, ST, or STP.
Timeouts
is a cumulative number that indicates how many times the module
has not responded to the monitor’s inquiry.
Bad checksums
is a cumulative
number that indicates how many times communication from the module to the
monitor has
broken down.
Bad c-s by mod
is a cumulative number that indicates how many
communication errors the module has detected.
The monitor starts counting these items at power up and resets to zero at
power off. The nonzero values do not indicate a failure, but the continuous
counting (more than 5 per second) indicates either a serial communication
failure, or a module not in place. Also other modules can cause communication
errors that cause these numbers rise.
RAM
indicates the state of the RAM memory.
ROM
indicates whether the checksum at the EPROM is in accordance with the
one the software has calculated.
EEPROM
indicates if the values stored in the permanent memory are valid.
The state is either
OK
,
Fail
or
?
(module not in place or a communication error).
Summary of Contents for b30
Page 2: ......
Page 12: ...B30 Patient Monitor 8 Document no 2044677 001...
Page 13: ...1 Introduction...
Page 14: ......
Page 61: ...2 Installation...
Page 69: ...3 Maintenance...
Page 70: ......
Page 105: ...4 Troubleshooting...
Page 106: ......
Page 125: ...5 Service Menu...
Page 126: ......
Page 158: ...B30 Patient Monitor 5 32 Document no 2044677 001...
Page 159: ...6 Field replaceable unit...
Page 160: ......
Page 166: ...B30 Patient Monitor 6 6 Document no 2044677 001 1 6 Frame for B30 Frame for B30...
Page 183: ...Frame disassembly 6 23 Document no 2044677 001...
Page 190: ...B30 Patient Monitor 6 30 Document no 2044677 001...
Page 191: ...7 Technical specification...
Page 192: ......
Page 198: ...B30 Patient Monitor 7 6 Document no 2044677 001...
Page 199: ...8 N F C REC Module...
Page 200: ......
Page 238: ...B30 Patient Monitor A 6 Document no 2044677 001...
Page 251: ......