E-PSM(P)W module introduction
1-33
Document no. 2044677-001
Figure 18
NIBP pneumatics diagram
Power supply section of the NIBP board
All connections are established via a 5-pin connector (female). The module needs a +15 V (dirty)
power supply to operate. The supply voltage Vmod 13.8- 16 V is generated in the power supply
section of the monitor. The other voltages needed for the operation of the NIBP measurement
are made on the NIBP board.
The NIBP power supply synchronizes the ECG and STP isolation power and supplies
non-isolated 5 V to the ECG and STP board.
4.4.3 ECG board in 5-lead measurement
The ECG measurement consists of the functions shown in
. All functions are located in
the ECG board except the ECG input unit.
Safety pressure sensor
Main pressure sensor
Dump valve
Proportional valve
Air pump
Intake air filter
Cuff connector
PSM
_N
IB
P
_p
neu
m_
di
ag
r.
vs
d
S1
S2
Plunger
Plunger
Summary of Contents for b30
Page 2: ......
Page 12: ...B30 Patient Monitor 8 Document no 2044677 001...
Page 13: ...1 Introduction...
Page 14: ......
Page 61: ...2 Installation...
Page 69: ...3 Maintenance...
Page 70: ......
Page 105: ...4 Troubleshooting...
Page 106: ......
Page 125: ...5 Service Menu...
Page 126: ......
Page 158: ...B30 Patient Monitor 5 32 Document no 2044677 001...
Page 159: ...6 Field replaceable unit...
Page 160: ......
Page 166: ...B30 Patient Monitor 6 6 Document no 2044677 001 1 6 Frame for B30 Frame for B30...
Page 183: ...Frame disassembly 6 23 Document no 2044677 001...
Page 190: ...B30 Patient Monitor 6 30 Document no 2044677 001...
Page 191: ...7 Technical specification...
Page 192: ......
Page 198: ...B30 Patient Monitor 7 6 Document no 2044677 001...
Page 199: ...8 N F C REC Module...
Page 200: ......
Page 238: ...B30 Patient Monitor A 6 Document no 2044677 001...
Page 251: ......