3.2 NIBP
3.2.1 NIBP toubleshooting
Problem
Cause
What to do
No NIBP value displayed
NIBP not selected on screen.
Check monitor setup.
‘Artifacts’ message
Unsuccessful measurement due to
patient movement, shivering,
external artifact or weak signal.
Check the patient status.
‘Weak pulsation’ message
Weak or unstable oscillation pulses
due to:
•
artifacts
•
weak pulse pressure due to
arrhythmias
•
improper cuff position or
attachment
•
too few pulses detected
•
weak or unusual blood
circulation
•
obese patient
Check patient condition and retry.
Check any leaks and retry.
Use proper size of cuff. Check
attachment.
Call service
‘Error X’ message
NIBP hardware error.
X = error number.
See the description of the error message
code.
‘Cuff loose’ message
1. Hose and/or cuff not connected. 1. Connect the hose and the cuff.
2. Hose and cuff connected.
Reasons:
- cuff loosely wrapped
- Tighten the cuff.
- leakage inside the shield, in the
Patient connector panel or
tubings connecting to the
module
- Check the tubings inside the shield
and Patient connector panel, fix if
necessary.
- leakage in cuff or hose
- Replace cuff/hose.
- leakage inside module
- Check internal tubing and fix if
necessary.
- pump does not work
- Check pump connector; if OK,
replace the NIBP Pump Unit.
Summary of Contents for b30
Page 2: ......
Page 12: ...B30 Patient Monitor 8 Document no 2044677 001...
Page 13: ...1 Introduction...
Page 14: ......
Page 61: ...2 Installation...
Page 69: ...3 Maintenance...
Page 70: ......
Page 105: ...4 Troubleshooting...
Page 106: ......
Page 125: ...5 Service Menu...
Page 126: ......
Page 158: ...B30 Patient Monitor 5 32 Document no 2044677 001...
Page 159: ...6 Field replaceable unit...
Page 160: ......
Page 166: ...B30 Patient Monitor 6 6 Document no 2044677 001 1 6 Frame for B30 Frame for B30...
Page 183: ...Frame disassembly 6 23 Document no 2044677 001...
Page 190: ...B30 Patient Monitor 6 30 Document no 2044677 001...
Page 191: ...7 Technical specification...
Page 192: ......
Page 198: ...B30 Patient Monitor 7 6 Document no 2044677 001...
Page 199: ...8 N F C REC Module...
Page 200: ......
Page 238: ...B30 Patient Monitor A 6 Document no 2044677 001...
Page 251: ......