![GE b30 Technical Reference Manual Download Page 129](http://html1.mh-extra.com/html/ge/b30/b30_technical-reference-manual_4013170129.webp)
Frame
5-3
Document no. 2044677-001
2
Frame
2.1 Memory
The Frame menu includes frame-specific service
menus.
A service menu to check the status of the
memory used in the CPU board of the monitor.
Test Memory
tests the condition of the
EEPROM/Flash memory component of the CPU
board. If the result of the test is
Fail
, see section
“Error messages” in chapter “CPU.”
Test SRAM
tests the Static RAM memory of the
CPU board in a similar way as the
EEPROM/Flash memory. If the result of the test
is
Fail
, see section “Error messages” in chapter
“CPU.”
Real-time clock
test is run at every start up and
also during the operation of the monitor. If the
result of the test is
Fail
, the battery for the SRAM
timekeeper should be replaced.
Summary of Contents for b30
Page 2: ......
Page 12: ...B30 Patient Monitor 8 Document no 2044677 001...
Page 13: ...1 Introduction...
Page 14: ......
Page 61: ...2 Installation...
Page 69: ...3 Maintenance...
Page 70: ......
Page 105: ...4 Troubleshooting...
Page 106: ......
Page 125: ...5 Service Menu...
Page 126: ......
Page 158: ...B30 Patient Monitor 5 32 Document no 2044677 001...
Page 159: ...6 Field replaceable unit...
Page 160: ......
Page 166: ...B30 Patient Monitor 6 6 Document no 2044677 001 1 6 Frame for B30 Frame for B30...
Page 183: ...Frame disassembly 6 23 Document no 2044677 001...
Page 190: ...B30 Patient Monitor 6 30 Document no 2044677 001...
Page 191: ...7 Technical specification...
Page 192: ......
Page 198: ...B30 Patient Monitor 7 6 Document no 2044677 001...
Page 199: ...8 N F C REC Module...
Page 200: ......
Page 238: ...B30 Patient Monitor A 6 Document no 2044677 001...
Page 251: ......