B30 Patient Monitor
5-24
Document no
.
2044677-001
4.4.3 NIBP Pulse Valve
How to check Pulse Valve
1.
Wrap an adult cuff around a pipe and connect the cuff to the module.
2.
Select
Start test
and push the ComWheel. The pressure rises beyond 240 mmHg and
stops. The pulse valve opens. The module counts the time it takes for the pressure to go
down from 240 mmHg to 50 mmHg and displays it on the screen.
3.
The test can be manually stopped by selecting
Stop test
.
4.
The valve can be adjusted between 0 and 255 (0 for fully closed and 255 for fully open).
First select
Set Valve
and push the ComWheel. See the pulse valve value and adjust it by
turning the ComWheel. Then push the ComWheel to confirm the value.
5.
The
Interval 240 mmHg -> 50 mmHg
time should be less than 60 seconds when the valve
is 150 and less than 10 when fully opened (255). When fully closed (0), the system should
be airtight and the pressure should not drop. Depending on the patient, the pulse valve
may remain closed up to approximately the value 100.
If the measured time deviates much from those above, the pulse valve or its tubes are faulty.
Start test
is for starting and
Stop test
is for
stopping the test.
Set Valve
lets you adjust the opening of the pulse
valve.
Pulse Valve Data
For information on general items
Pressure
,
Zero
,
Protect handle
,
Calibr. prot.
,
+15 V power
,
AD0
to
AD7
as well as
Timeouts
etc.,
Summary of Contents for b30
Page 2: ......
Page 12: ...B30 Patient Monitor 8 Document no 2044677 001...
Page 13: ...1 Introduction...
Page 14: ......
Page 61: ...2 Installation...
Page 69: ...3 Maintenance...
Page 70: ......
Page 105: ...4 Troubleshooting...
Page 106: ......
Page 125: ...5 Service Menu...
Page 126: ......
Page 158: ...B30 Patient Monitor 5 32 Document no 2044677 001...
Page 159: ...6 Field replaceable unit...
Page 160: ......
Page 166: ...B30 Patient Monitor 6 6 Document no 2044677 001 1 6 Frame for B30 Frame for B30...
Page 183: ...Frame disassembly 6 23 Document no 2044677 001...
Page 190: ...B30 Patient Monitor 6 30 Document no 2044677 001...
Page 191: ...7 Technical specification...
Page 192: ......
Page 198: ...B30 Patient Monitor 7 6 Document no 2044677 001...
Page 199: ...8 N F C REC Module...
Page 200: ......
Page 238: ...B30 Patient Monitor A 6 Document no 2044677 001...
Page 251: ......