B30 Patient Monitor
8-16
Document no
.
2044677-001
Recorder
17. Paper compartment cleaning
Open the paper compartment hatch and take out the paper roll, if installed.
Remove any paper chaff from the compartment.
Clean the thermal printhead and the small glass window in front of the static brush with a
cotton swab dipped in isopropyl alcohol. Avoid contact with the rubber paper roller.
NOTE: Be careful to limit the application of alcohol to the thermal printhead and the window.
Leave the paper compartment empty and close the hatch.
"
Configure the monitor screen so that all required parameters are shown, for example:
Monitor Setup
- Screen 1 Setup - Waveform Fields - Field 1 - ECG1
Field 2 - ECG2
Field 3 - P1
Field 4 - P2
Field 5 - PLETH
Field 6 - OFF
Plug in the Hemodynamic Module. Connect a patient simulator to the module and check
that all connected parameters are shown on the screen.
Preset recording settings:
Record/Print
-
Record Waveforms
-
Waveform 1 - ECG1
Waveform 2 - P1
Waveform 3 - P2
Delay - Off
Paper Speed - 6.25 Mm/S
Length --> 30 S
Record Trends - Graphic Trend 1 - Hr
Graphic Trend 2 - P1
Monitor Setup
-
Install/Service
(password 16-4-34)
- Installation - Printer & Recorder -
Default Trend - Graph.
Display Trends - Time Scale - 2 h
18. Paper recognition
Press the
Start/Stop
key on the monitor side panel. Check that the message “Recorder:
Out of paper’ appears on the screen.
"
19. Cover state recognition
Open the paper compartment cover. Check that the previous message changes to
‘Recorder: Cover open’.
Install a paper roll and close the cover. Check that the message ‘Recorder: Cover open’
disappears from the screen.
"
Summary of Contents for b30
Page 2: ......
Page 12: ...B30 Patient Monitor 8 Document no 2044677 001...
Page 13: ...1 Introduction...
Page 14: ......
Page 61: ...2 Installation...
Page 69: ...3 Maintenance...
Page 70: ......
Page 105: ...4 Troubleshooting...
Page 106: ......
Page 125: ...5 Service Menu...
Page 126: ......
Page 158: ...B30 Patient Monitor 5 32 Document no 2044677 001...
Page 159: ...6 Field replaceable unit...
Page 160: ......
Page 166: ...B30 Patient Monitor 6 6 Document no 2044677 001 1 6 Frame for B30 Frame for B30...
Page 183: ...Frame disassembly 6 23 Document no 2044677 001...
Page 190: ...B30 Patient Monitor 6 30 Document no 2044677 001...
Page 191: ...7 Technical specification...
Page 192: ......
Page 198: ...B30 Patient Monitor 7 6 Document no 2044677 001...
Page 199: ...8 N F C REC Module...
Page 200: ......
Page 238: ...B30 Patient Monitor A 6 Document no 2044677 001...
Page 251: ......