![GE b30 Technical Reference Manual Download Page 219](http://html1.mh-extra.com/html/ge/b30/b30_technical-reference-manual_4013170219.webp)
N-FCREC Module disassembly and reassembly
8-19
Document no. 2044677-001
3. Unscrew the two pozidrive screws inside the
recorder unit completely (the screws cannot be
removed) and remove the recorder unit.
4. Remove the five screws (T8) on the side where the
device plate is located.
5. Place the N-F(C)(REC) on a table the device plate
side down.
6. Remove the two screws (T8) near the module bus
connector.
7. Remove the module cover carefully.
8. Remove the EMC-cover and CO
2
unit insulator
plate.