![GE b30 Technical Reference Manual Download Page 141](http://html1.mh-extra.com/html/ge/b30/b30_technical-reference-manual_4013170141.webp)
Parameters
5-15
Document no. 2044677-001
4.2 ECG Module
Service Data
Power Freq
shows the mains frequency selected: 50 Hz
/ 60 Hz
Cable type
shows the leadwire set connected: 3 lead /
5 lead
Quick Zero
shows 1 when the ECG signal is beyond
scale, and therefore, is quickly returned to the optimal
range with fast signal processing methods.
Artifact
shows 1 when artifacts are detected.
Electrode
shows ON when each of these electrodes is
connected.
Pacer count
is a running number of detected
pacemaker spikes.
R count
is a running number of detected R waves.
Resp: Avail
shows if a module with impedance
measurement is plugged (1) or not plugged (0) into the
monitor.
Value
shows the measured respiration rate value for impedance respiration.
Zeroing
indicates the zeroing status of the respiration measurement: 1 = zeroing, 0 = not zeroing.
MeasOff
shows 1 if the respiration measurement is set to OFF, and 0 if the the respiration
measurement is set to ON.
Arrhythmia
shows an active arrhythmia alarm (VFIB, VTAC, ASY).
On/Off
shows 1 if arrhythmia analysis is active and 0 if it is not active.
HR, HR25
and
HR75
for R&D purpose only
QRS
count shows the running number of detected QRS complexes.
PVC
shows the PVCs detected per minute
Det leads
shows the leads that are being used for detecting beats and ventricular fibrillation. The
selection of user leads (ECG1, ECG2, ECG3) on the monitor affects the leads used for detection. The first
lead used for detection is lead either I or II. The algorithm uses the lead appearing first in user leads. The
second lead used for detection is one of the precordial leads (V1 - V6): -1 = invalid (not used), 0 = lead I,
1 = II or III; 2 = V1, 3 = V2, 4 = V3, 5 = V4, 6 =V5, 7 = V6
nroleads
shows the number of leads that are beign used for detecting beats and ventricular fibrillation.
Possible values include: 0 , 1 and 2 (ref.
Det leads
) analysis.
Data
for R&D purpose only
Noise M/A
for R&D purpose only
Timeouts
is a cumulative number that indicates how many times the module has not responded to the
monitor’s inquiry.
Bad checksums
is a cumulative number that indicates how many times
communication from the module to the monitor has broken down.
Summary of Contents for b30
Page 2: ......
Page 12: ...B30 Patient Monitor 8 Document no 2044677 001...
Page 13: ...1 Introduction...
Page 14: ......
Page 61: ...2 Installation...
Page 69: ...3 Maintenance...
Page 70: ......
Page 105: ...4 Troubleshooting...
Page 106: ......
Page 125: ...5 Service Menu...
Page 126: ......
Page 158: ...B30 Patient Monitor 5 32 Document no 2044677 001...
Page 159: ...6 Field replaceable unit...
Page 160: ......
Page 166: ...B30 Patient Monitor 6 6 Document no 2044677 001 1 6 Frame for B30 Frame for B30...
Page 183: ...Frame disassembly 6 23 Document no 2044677 001...
Page 190: ...B30 Patient Monitor 6 30 Document no 2044677 001...
Page 191: ...7 Technical specification...
Page 192: ......
Page 198: ...B30 Patient Monitor 7 6 Document no 2044677 001...
Page 199: ...8 N F C REC Module...
Page 200: ......
Page 238: ...B30 Patient Monitor A 6 Document no 2044677 001...
Page 251: ......