Parameters
5-17
Document no. 2044677-001
4.3 STP Module
Record Data
prints out the shown service
data and board information (id, serial number
and sw id) onto the recorder.
Temp Test
activates the automatic
temperature test for the temperature
channels T1 and T2. The result from the test is
shown in the service data field.
NOTE: The Temp Test needs to be selected
twice before the test starts.
Service Data field
Gain
is a coefficient to compensate gain
error. Usually the values for P1 and P2 are
between 17000 and 25000 and for T1 and T2
between 13000 and 14300.
Zero
indicates
the offset compensation value of each
parameter in the A/D converter. Typically the
values for P1 and P2 are within
±
1000 and for
T1 and T2 between -150 and +300. Calibrate
if zero and/or gain value is outside the
ranges.
Cable
shows ON when a corresponding cable
is connected to the front panel and
Probe
shows ON when a corresponding probe is connected to the cable.
Under
Value
,
the measured numeric values are displayed simultaneously. Pressure values are real time
values and shown in mmHg. Temperature values are shown in degrees Celsius.
The front panel STP keys functions are confirmed by pressing each key and observing that
OFF
turns to
ON
at
Button
.
SpO2
shows the measured beat-to-beat SpO
2
value.
Modpr
is a modulation % that indicates the AC/DC
ratio in the measured signal.
Hr
is a pulse rate calculated from every beat.
Cable
and
Probe
can be either
OFF
or
ON,
and these indicate the state PROBE OFF.
Under them there is a message field for SpO
2
. It can be OK, PULSE SEARCH, NO PROBE, PROBE OFF, NO
PULSE, ARTEFACT, POOR SIGNAL, or CHECK PROBE.
Balance between leds is adjusted by changing the intensity of red/infrared. Intensity of infrared (
Ired
Int.
) is in the range of 40 to 255 and red intensity (
Red Int.
) is in the range of 40 to 255.
DC gain
shows the gain of DC signal adjusted by the module.
IDC
is the value of infrared signal.
RDC
is the dc value of red signal.
AC gain
is the gain of infrared and red ac signals. AC gain values can be 1 or 0. Value 1 means high ac
gain and 0 means low gain.
Pre gain
is a preamplifier gain for infrared and red signals. Pre gain values can be 1 or 0. Value 1 means
normal operation. Value 0 means that signal levels are very low and extra gain is taken into use.
Summary of Contents for b30
Page 2: ......
Page 12: ...B30 Patient Monitor 8 Document no 2044677 001...
Page 13: ...1 Introduction...
Page 14: ......
Page 61: ...2 Installation...
Page 69: ...3 Maintenance...
Page 70: ......
Page 105: ...4 Troubleshooting...
Page 106: ......
Page 125: ...5 Service Menu...
Page 126: ......
Page 158: ...B30 Patient Monitor 5 32 Document no 2044677 001...
Page 159: ...6 Field replaceable unit...
Page 160: ......
Page 166: ...B30 Patient Monitor 6 6 Document no 2044677 001 1 6 Frame for B30 Frame for B30...
Page 183: ...Frame disassembly 6 23 Document no 2044677 001...
Page 190: ...B30 Patient Monitor 6 30 Document no 2044677 001...
Page 191: ...7 Technical specification...
Page 192: ......
Page 198: ...B30 Patient Monitor 7 6 Document no 2044677 001...
Page 199: ...8 N F C REC Module...
Page 200: ......
Page 238: ...B30 Patient Monitor A 6 Document no 2044677 001...
Page 251: ......