N-FCREC Module disassembly and reassembly
8-21
Document no. 2044677-001
11. Remove the CO2 unit module bus flex cable from
the REC board.
12. Lift the CO2 unit up carefully, disconnect the
module bus flex cable and remove the unit from
the module cover.
13. Remove the screw (T8) holding the module bus flex
cable to recorder unit metal plate and module
cover.
14. Disconnect the module bus flex cable from the REC
board and remove the REC board with the recorder
unit metal plate from the module cover.
15. Remove the four pozidrive screws holding the REC
board to the recorder unit metal plate.
Summary of Contents for b30
Page 2: ......
Page 12: ...B30 Patient Monitor 8 Document no 2044677 001...
Page 13: ...1 Introduction...
Page 14: ......
Page 61: ...2 Installation...
Page 69: ...3 Maintenance...
Page 70: ......
Page 105: ...4 Troubleshooting...
Page 106: ......
Page 125: ...5 Service Menu...
Page 126: ......
Page 158: ...B30 Patient Monitor 5 32 Document no 2044677 001...
Page 159: ...6 Field replaceable unit...
Page 160: ......
Page 166: ...B30 Patient Monitor 6 6 Document no 2044677 001 1 6 Frame for B30 Frame for B30...
Page 183: ...Frame disassembly 6 23 Document no 2044677 001...
Page 190: ...B30 Patient Monitor 6 30 Document no 2044677 001...
Page 191: ...7 Technical specification...
Page 192: ......
Page 198: ...B30 Patient Monitor 7 6 Document no 2044677 001...
Page 199: ...8 N F C REC Module...
Page 200: ......
Page 238: ...B30 Patient Monitor A 6 Document no 2044677 001...
Page 251: ......