Service procedures
8-17
Document no. 2044677-001
20. Quality of recording
Press the
Start/ Stop
key again and check that the quality of the recordings is
acceptable.
"
21. Recording speed
Press the
RECORD WAVE
key again and this time wait until the recording stops. Check
that the length of the recorded waveform scale is 18.7 cm (
±
1.5 cm).
Change the paper speed setting to 1 mm/s:
Record/Print
- Record Waveforms - Paper Speed - 1 mm/s
Press the
RECORD WAVE
key and wait until the recording stops. Check that the length
of the scale is now 3.0 cm (
±
0.5 cm).
"
General
22. Electrical safety check
Perform an electrical safety check and a leakage current test.
"
23. Functioning after electrical safety check
Check that the module functions normally after performing the electrical safety check.
"
24. Final cleaning
Turn off the monitor, disconnect and clean the module.
"
•
Fill in all necessary documents.
Summary of Contents for b30
Page 2: ......
Page 12: ...B30 Patient Monitor 8 Document no 2044677 001...
Page 13: ...1 Introduction...
Page 14: ......
Page 61: ...2 Installation...
Page 69: ...3 Maintenance...
Page 70: ......
Page 105: ...4 Troubleshooting...
Page 106: ......
Page 125: ...5 Service Menu...
Page 126: ......
Page 158: ...B30 Patient Monitor 5 32 Document no 2044677 001...
Page 159: ...6 Field replaceable unit...
Page 160: ......
Page 166: ...B30 Patient Monitor 6 6 Document no 2044677 001 1 6 Frame for B30 Frame for B30...
Page 183: ...Frame disassembly 6 23 Document no 2044677 001...
Page 190: ...B30 Patient Monitor 6 30 Document no 2044677 001...
Page 191: ...7 Technical specification...
Page 192: ......
Page 198: ...B30 Patient Monitor 7 6 Document no 2044677 001...
Page 199: ...8 N F C REC Module...
Page 200: ......
Page 238: ...B30 Patient Monitor A 6 Document no 2044677 001...
Page 251: ......