E-PSM(P)W module disassembly
6-25
Document no. 2044677-001
- Disconnect the module bus connector, pump
connector and NIBP flex connector.
- Disconnect the hoses (2 pcs) coming from the
manifold.
- Lift up the NIBP board.
NOTE: Note the positions of the hoses; mark them if
necessary to ensure they are replaced correctly.
To remove the NIBP pump and the
manifold unit
- Disconnect the air intake hose from the NIBP manifold.
- Remove the NIBP filter.