Electrical Safety Tests
3-7
Document no. 2044677-001
•
Neutral - OPEN
•
GND (Earth) - CLOSED
(5) Read and record the current leakage indicated on the tester.
(6) Change leakage tester switches to:
•
Polarity - NORMAL
•
Neutral - OPEN
•
GND (Earth) - CLOSED
(7) Read and record the current leakage indicated on the tester.
(8) Change leakage tester switches to:
•
Polarity - NORMAL
•
Neutral - OPEN
•
GND (Earth) - CLOSED
(9) Read and record the current leakage indicated on the tester.
If measured reading is greater than the appropriate specification below, the device under test
fails. Contact GE Technical Support.
2.7.1 Acceptance criteria NC (Normal condition)
•
(USA only) 300 µA, and the device under test is powered from 100-120 V/50-60 Hz
•
(USA only) 300 µA, and the device under test is powered from a center-tapped 200-240
V/50-60 Hz, single phase circuit (UL Split Phase Exemption)
•
500 µA, and the device under test is powered from a non-centertapped, 200-240 V/50-60
Hz, single-phase circuit
2.7.2 Acceptance criteria SFC (Single fault condition) - ground (earth), line or
neutral open
•
(USA only) 300 µA, and the device under test is powered from 100-120 V/ 50-60 Hz
•
(USA only) 300 µA, and the device under test is powered from a center-tapped 200-240
V/50-60 Hz, single phase circuit (UL Split Phase Exemption)
•
1000 µA
NOTE: Center-tapped and non-center-tapped supply circuits produce different leakage
currents and the UL and IEC limits are different.
Summary of Contents for b30
Page 2: ......
Page 12: ...B30 Patient Monitor 8 Document no 2044677 001...
Page 13: ...1 Introduction...
Page 14: ......
Page 61: ...2 Installation...
Page 69: ...3 Maintenance...
Page 70: ......
Page 105: ...4 Troubleshooting...
Page 106: ......
Page 125: ...5 Service Menu...
Page 126: ......
Page 158: ...B30 Patient Monitor 5 32 Document no 2044677 001...
Page 159: ...6 Field replaceable unit...
Page 160: ......
Page 166: ...B30 Patient Monitor 6 6 Document no 2044677 001 1 6 Frame for B30 Frame for B30...
Page 183: ...Frame disassembly 6 23 Document no 2044677 001...
Page 190: ...B30 Patient Monitor 6 30 Document no 2044677 001...
Page 191: ...7 Technical specification...
Page 192: ......
Page 198: ...B30 Patient Monitor 7 6 Document no 2044677 001...
Page 199: ...8 N F C REC Module...
Page 200: ......
Page 238: ...B30 Patient Monitor A 6 Document no 2044677 001...
Page 251: ......