E-PSM(P)W Troubleshooting
4-13
Document no. 2044677-001
3.2.2 NIBP error code explanation
Code
Problem
What to do
0
RAM failure; memory failure
Change the NIBP board.
1
ROM checksum error; memory failure
Change the NIBP board.
2
Pump current failure
Check short circuits. Change the NIBP board.
3
Safety CPU internal test failure or pressure
sensor reference voltage failure
Change the NIBP board.
4
Calibration switch is ON
Press module buttons
Auto ON/OFF
and
Start/
Cancel
simultaneously for 3 seconds.
5
Calibration not protected
Protect calibration by selecting Protection ON in
the NIBP calibration menu.
6
Pressure sensors give different readings
Try to remeasure. If the problem persists,
recalibrate. If the problem still persists, change
the NIPB board.
7
Calibration failure
Reset the module and recalibrate. If this does not
help, change the NIBP board.
8
Exhaust Valve occlusion
Check and clean the tubing and air chamber. If
this does not help, change the NIBP board.
9
Measurement related error
Automatic recovery.
10
EEPROM checksum error; memory failure
Change the NIBP board.
11
Auto zero range exceeded
Calibrate the NIBP.
12
Not in use
Not in use
13
Illegal neonate cuff with identifying magnet
connected
Remove the cuff.
14
Not in use
Not in use
15
Safety CPU pressure calibration error
Recalibrate. If this does not help, change the
NIBP board.
16
Communication error between CPUs
Change the NIBP board.
17
Safety CPU has cut down power from
pneumatics due to repeating safety limit
violations
Reset the module. If the problem persists, change
the NIBP board.
Summary of Contents for b30
Page 2: ......
Page 12: ...B30 Patient Monitor 8 Document no 2044677 001...
Page 13: ...1 Introduction...
Page 14: ......
Page 61: ...2 Installation...
Page 69: ...3 Maintenance...
Page 70: ......
Page 105: ...4 Troubleshooting...
Page 106: ......
Page 125: ...5 Service Menu...
Page 126: ......
Page 158: ...B30 Patient Monitor 5 32 Document no 2044677 001...
Page 159: ...6 Field replaceable unit...
Page 160: ......
Page 166: ...B30 Patient Monitor 6 6 Document no 2044677 001 1 6 Frame for B30 Frame for B30...
Page 183: ...Frame disassembly 6 23 Document no 2044677 001...
Page 190: ...B30 Patient Monitor 6 30 Document no 2044677 001...
Page 191: ...7 Technical specification...
Page 192: ......
Page 198: ...B30 Patient Monitor 7 6 Document no 2044677 001...
Page 199: ...8 N F C REC Module...
Page 200: ......
Page 238: ...B30 Patient Monitor A 6 Document no 2044677 001...
Page 251: ......