B30 Patient Monitor
3-10
Document no
.
2044677-001
2.9 Patient (source) leakage current test
This procedure only applies to Class I (grounded/earthed) equipment, and measures the
leakage current from the ECG/RESP connector or the SPO2 connector of the device to ground.
NOTE: The MD represents the network and voltage measuring instrument and its frequency
characteristics per IEC 60601-1.
The patient connector test body shorts all signals in the connector together. Refer to the
instructions contained with the safety analyzer to perform this test.
(1) Connect the ECG/RESP Test Body to the green connector of the device under test.
(2) Configure leakage tester as follows:
•
Polarity - NORMAL
•
Neutral - OPEN
•
GND (Earth) - CLOSED
(3) Power on Device under test.
(4) Read and record the current leakage indicated on the tester.
(5) Change leakage tester switches to:
•
Polarity - NORMAL
•
Neutral - OPEN
•
GND (Earth) - CLOSED
(6) Read and record the current leakage indicated on the tester.
(7) Change leakage tester switches to:
•
Polarity - NORMAL
•
Neutral - OPEN
•
GND (Earth) - CLOSED
(8) Read and record the current leakage indicated on the tester.
(9) Change leakage tester switches to:
•
Polarity - NORMAL
•
Neutral - OPEN
Power Cord
Leakage Tester
NORM
RVS
ECG Test Body
or SPO2 Test Body
HIGH
LOW
GND
1K
0.15µF
10
Power Cord
Device
Under
Test
DMM set to measure AC voltage
GND
DMM
Closed
Summary of Contents for b30
Page 2: ......
Page 12: ...B30 Patient Monitor 8 Document no 2044677 001...
Page 13: ...1 Introduction...
Page 14: ......
Page 61: ...2 Installation...
Page 69: ...3 Maintenance...
Page 70: ......
Page 105: ...4 Troubleshooting...
Page 106: ......
Page 125: ...5 Service Menu...
Page 126: ......
Page 158: ...B30 Patient Monitor 5 32 Document no 2044677 001...
Page 159: ...6 Field replaceable unit...
Page 160: ......
Page 166: ...B30 Patient Monitor 6 6 Document no 2044677 001 1 6 Frame for B30 Frame for B30...
Page 183: ...Frame disassembly 6 23 Document no 2044677 001...
Page 190: ...B30 Patient Monitor 6 30 Document no 2044677 001...
Page 191: ...7 Technical specification...
Page 192: ......
Page 198: ...B30 Patient Monitor 7 6 Document no 2044677 001...
Page 199: ...8 N F C REC Module...
Page 200: ......
Page 238: ...B30 Patient Monitor A 6 Document no 2044677 001...
Page 251: ......