Equipotential Grounding
When connecting multiple equipments, electrical potential difference may be
generated between the equipments. This may result in electric shock to the
patient connected to these equipments. Pay special attention for use in
Operating Room, ICU, CCU, Cardiac Catheter Laboratory, and Cardiovascular
X-ray room. To avoid such electrical potential difference, use the ground cable to
connect each device’s potential equalization terminal to the same ground
terminal. This is called equipotential grounding.
The ground cable is shaped like a
connector, and can be disconnected
manually without using tools.