CAUTION
Systems
The DSL-8001 is intended for patient monitoring in ICU,
CCU, surgery, ward and during transportation within the
hospital. Do not use the monitor out of the hospital (like in
airplane or ambulance car), in MRI environment, inside
hyperbaric chamber, or in the presence of flammable
anaesthetics.
The use of alarm output function needs to be tested after it
has been set up in your facility. It should be tested
whenever setting up the DSL-8001 system in a location
that uses the alarm system. One way to test alarm output
function is to create an alarm condition and verify that your
facility’s alarm system is activated.
Use only the accessories specified for this device.
Otherwise, proper function cannot be executed.
For quality improvement, specifications are subject to
change without prior notice.
This equipment is intended to be used for only one patient.
The operation and maintenance of this equipment should
be performed by well-trained and authorized personnel.
The installation of this equipment should be performed by
our service representative or a person who is well
acquainted with this equipment.
The installation of the battery will be performed by our
service representative. Users should not perform this
procedure as electric shock may result to the patient and
operator or malfunction of the equipment may occur.
If not using for a long period, make sure to turn OFF the
power of the main unit.
When lifting this equipment, hold it by the handle or the
bottom part of the main unit. During transportation, make
sure to firmly grasp the handle of the monitor to avoid
dropping it. The operator may be injured or the equipment
may be damaged.
When installing the recording paper, pay attention not to
touch the thermal head or sensor. The temperature of
those parts rises immediately after printing and may cause
burn injury. Also, it may cause failure to the thermal head
and sensor.
To turn OFF the power of this unit, press the power switch
on the front side. Do not unplug the AC power cable before
the display is turned OFF.
Patient Admit / Discharge
This unit cannot display the patient information (patient ID,
patient name, etc.) received from the central monitor, even
when a wireless network system is constructed.
The patient classification selection influences the precision
of the QRS detection or NIBP measurement and Alarm
Limit Setup area. Make sure the correct selection is made.
If monitoring of new patient is started without discharging
the previous patient, the measurement data of the
previous and new patient will become mixed up on the
table and trend data.
Summary of Contents for DSL-8001
Page 1: ......
Page 2: ......
Page 34: ...xxxii ...
Page 50: ...1 8 ...
Page 56: ...2 6 Setup 2 2 window ...
Page 88: ...2 38 Reference For details refer to 4 Monitoring Setup Printing Setup ...
Page 90: ...2 40 ...
Page 112: ...3 22 ...
Page 166: ...6 10 ...
Page 182: ...7 16 ...
Page 216: ...8 34 ...
Page 250: ...10 6 ...
Page 251: ......
Page 252: ......