Check Discharge at Power ON
The discharge operation will be automatically performed when the power is
turned ON again after the power was turned OFF for preprogrammed duration.
Whether or not to display the confirmation window for the discharge operation
can be selected.
1 Press the
(Menu) key, and select “Maintenance” > enter the
password > “Backup Setup” > “Check Discharge at Power ON”.
2 Select
[ON]/[OFF].
If [ON] is selected, the confirmation window to perform the discharge operation
will be displayed at power ON if the power OFF duration exceeds the set duration
for “Automatic Discharge Time” (10 min. / 1 hour / 2 hours / 4 hours).
If [OFF] is selected, the confirmation window will not be displayed and the
discharge operation will be automatically performed at power ON if the power
OFF duration exceeds the set duration for “Automatic Discharge Time”.
However, if “Automatic Discharge Time” is set to [OFF] and “Check Discharge at
Power ON” is set to [ON], the confirmation window will be displayed at power ON
if the power OFF duration exceeds 30 seconds.