8
-
31
The <Printer Communication Error> message is displayed.
Cause
: Communication error with the recorder board has occurred.
Solution : Reboot the equipment. If the same message is repeatedly displayed,
a failure of the equipment can be considered. Contact our service
representative.
Telemeter
The <HLX Module Error> message is displayed.
Cause
: There is a communication error with the telemetry unit.
Solution : A broken wire or failure of the telemetry unit can be considered.
Contact our service representative.
There is no reception at the telemetry center.
Cause
: The channel ID or group ID does not match the telemetry receiver.
Solution : Set the correct channel ID and group ID.
The <HLX Invalid Setting> message is displayed.
Cause
: The telemetry channel setting is outside the allowable range.
Solution : Make sure the channel number is within the allowable range on
“Initial Setup” > “Telemeter Setup” > “Channel”.
The <HLX Data Saving Error> message is displayed.
Cause
: The set value cannot be written to HLX-801.
Solution : Reboot the equipment. If the same message is repeatedly displayed,
a failure of the equipment can be considered. Contact our service
representative.
The <HLX Check Connection> message is displayed.
Cause
: The connection with the HLX-801 is interrupted.
Solution : Check the connection between the HLX-801 and this equipment.
Make sure that [HLX] is set for “Telemeter Port” (“Initial Setup” >
“Communication Setup”).
Ventilator
The <Vent. Alarm> message is displayed.
Cause
: The following alarm has generated on the ventilator.
・
Parameter alarm such as AWP, MV, FiO
2
・
Technical alarm such as cable disconnected condition.
Solution : Check the alarm cause of the ventilator, and take appropriate action.
The <Vent. Invalid> message is displayed.
Cause
: The connection cable with the ventilator is not properly connected.
Solution : Make sure the connection cable with the ventilator is properly
connected.
The <Vent. Disable> message is displayed.
Cause
: [OFF] is set for “VENT Alarm” (“Initial Setup” > “VENT Setup”).
Summary of Contents for DSL-8001
Page 1: ......
Page 2: ......
Page 34: ...xxxii ...
Page 50: ...1 8 ...
Page 56: ...2 6 Setup 2 2 window ...
Page 88: ...2 38 Reference For details refer to 4 Monitoring Setup Printing Setup ...
Page 90: ...2 40 ...
Page 112: ...3 22 ...
Page 166: ...6 10 ...
Page 182: ...7 16 ...
Page 216: ...8 34 ...
Page 250: ...10 6 ...
Page 251: ......
Page 252: ......