Monitor Suspend at Discharge
Whether or not to suspend the monitoring when a patient is discharged can be
selected.
1 Press the
(Menu) key, and select “Maintenance” > enter the
password > “Backup Setup” > “Monitor Suspend at Discharge”.
2 Select
[ON]/[OFF].
If [ON] is selected, monitoring will be suspended after the discharge operation.
The numeric data display will be cleared, and alarm generation, NIBP periodic
measurement, periodic printing will not be performed.
If [OFF] is selected, monitoring will continue even after a patient is discharged.
●
Measurement Interval at Discharge
NIBP measurement interval after a patient is discharged can be selected.
Even if NIBP measurement interval is changed for each patient, this setting will
be applied after discharge.
1 Select the
(Menu) key, and select “Maintenance” > enter the
password > “Backup Setup” > “Measurement Interval at Discharge”.
The “Measurement
Interval at Discharge”
window will be
displayed.
2 Select the NIBP measurement interval at discharge.
(OFF / 2 min. / 2.5 min. / 5 min. / 10 min. / 15 min. / 30 min. / 60 min. / 120
min.)
Selecting [OFF] will turn off the NIBP auto mode after the discharge procedure.