xxix
Compliance to the Electromagnetic Immunity (1)
The DSL-8001 system is intended for use in the electromagnetic environment
specified below. It is recommended that the DSL-8001 system be used in such
an environment.
Immunity
Test
IEC60601-1-2
Test Level
Compliance Level
Electromagnetic
Environment -
Guidance
Electrostatic
Discharge (ESD)
IEC 61000-4-2
2, 4, 6kV Contact
2, 4, 8kV Air
2, 4, 6kV Contact
2, 4, 8kV Air
Floors should be
wood, concrete or
ceramic tile. If floors
are covered with
synthetic material, the
relative humidity
should be at least
30%.
Electrical fast
transient/burst
IEC 61000-4-4
2kV for power
supply lines
1kV for input/output
lines
2kV for power
supply lines
1kV for input/output
lines
Mains power quality
should be that of a
typical commercial or
hospital environment.
Surge
IEC 61000-4-5
0.5, 1kV:
differential mode
0.5, 1, 2kV:
common mode
0.5, 1kV:
differential mode
0.5, 1, 2kV:
common mode
Mains power quality
should be that of a
typical commercial or
hospital environment.
Voltage dips, short
interruptions, and
voltage variations
on power supply
input lines
IEC 61000-4-11
for 0.5 cycles
<5% U
T
(>95% dip in U
T
)
for 5 cycles
40% U
T
(60% dip in U
T
)
for 25 cycles
70% U
T
(30% dip in U
T
)
for 5 cycles
<5% U
T
(95% dip in U
T
)
for 0.5 cycles
<5% U
T
(>95% dip in U
T
)
for 5 cycles
40% U
T
(60% dip in U
T
)
for 25 cycles
70% U
T
(30% dip in U
T
)
for 5 cycles
<5% U
T
(95% dip in U
T
)
Mains power quality
should be that of a
typical commercial or
hospital environment.
If the user of the
DSL-8001 requires
continued operation
during power mains
interruptions, it is
recommended that the
DSL-8001 is equipped
with an internal battery
(option) or is powered
from an uninterruptible
power supply.
Power Frequency
(50/60Hz)
Magnetic Field
IEC 61000-4-8
3A/m 3A/m Power
frequency
magnetic fields should
be at levels
characteristic of a
typical location in a
typical commercial or
hospital environment.
Notes: U
T
is the AC mains voltage prior to application of the test level.
Summary of Contents for DSL-8001
Page 1: ......
Page 2: ......
Page 34: ...xxxii ...
Page 50: ...1 8 ...
Page 56: ...2 6 Setup 2 2 window ...
Page 88: ...2 38 Reference For details refer to 4 Monitoring Setup Printing Setup ...
Page 90: ...2 40 ...
Page 112: ...3 22 ...
Page 166: ...6 10 ...
Page 182: ...7 16 ...
Page 216: ...8 34 ...
Page 250: ...10 6 ...
Page 251: ......
Page 252: ......