2
-
32
To resume the alarm sound for all parameters
The alarm silence state will cease in the event of any of the following.
・
When the power is turned ON.
・
When the patient is discharged.
・
When the
(Alarm Silence) key is pressed.
To resume the alarm sound for each parameter
The alarm silence state will cease when the alarm of each parameter is set to
OFF.
●
To Suspend the Alarm Sound
The "Alarm Sound Suspend" function allows to suspend the alarm occurrence in
advance at a time such as during operation when the alarm generation is
expected.
To suspend the alarm sound, press the
(Alarm Silence) key for more than
3 seconds. During the alarm sound suspend state, the
(Alarm Silence)
icon will be displayed.
Alarm monitoring continues during the alarm sound suspend state. The alarm
sound suspend duration can be selected from 60 min. / 90 min. /120 min.
The alarm printing will be performed even during the alarm sound suspend state.
The “Alarm Sound Suspend Reminder” can be set to notify the alarm sound
suspend state with the selected interval (OFF/ 3 min./ 10 min.).
The alarm sound suspend state will cease in the event of any of the following.
・
When the patient is discharged.
・
When the
(Alarm Silence) key is pressed for 3 seconds or more.
Summary of Contents for DSL-8001
Page 1: ......
Page 2: ......
Page 34: ...xxxii ...
Page 50: ...1 8 ...
Page 56: ...2 6 Setup 2 2 window ...
Page 88: ...2 38 Reference For details refer to 4 Monitoring Setup Printing Setup ...
Page 90: ...2 40 ...
Page 112: ...3 22 ...
Page 166: ...6 10 ...
Page 182: ...7 16 ...
Page 216: ...8 34 ...
Page 250: ...10 6 ...
Page 251: ......
Page 252: ......