5
-
15
NIBP Data Display Time
Select the numeric data display duration (60min. / 120min.) inside the NIBP
numeric data box.
The data will be deleted after the selected duration.
1 Press the
(Menu) key, and select “Parameter Setup” > “NIBP Setup
(2/2) > “NIBP Data Display Time”.
2 Select [60 min.] or [120 min.].
NIBP Alarm
Select ON/OFF of NIBP alarm, and set the alarm limit.
1 Press the
(Menu) key and select “Parameter Setup” > “NIBP Setup
(2/2)” > “NIBP Alarm”.
2 Select [ON] or [OFF].
3 Next, set the alarm limit.
Select the "Alarm Limit Setup" to display the “Alarm Limit Setup” window.
(1) Select a parameter for setup.
(2) Select [ON] / [OFF] for the alarm.
If [OFF] is selected, alarm will be turned OFF and
(Alarm OFF) icon will
be displayed.
(3) Aligning the highlight and pressing the jog dial will change the numeric color to
yellow enabling selection of value by turning the jog dial. Press the jog dial to
finalize the setup.
(3)
(1)
(2)
Summary of Contents for DSL-8001