9
-
5
Ventricular Tachycardia 2mVpp, 206bpm
: Average 6.3 sec.
Ventricular Tachycardia 0.5mVpp, 206bpm : Average 13.3 sec.
Ventricular Tachycardia 2mVpp, 195bpm
: Average 7.8 sec.
Ventricular Tachycardia 4mVpp, 195bpm
: Average 8.4 sec.
Ventricular Tachycardia 1mVpp, 195bpm
: Average 7.9 sec.
Respiration
Method
: Impedance Method
Frequency Characteristic : 2.5Hz
Current :
100
A or lower
Measurement Range
: 0, 3 to 150Bpm
Measurement Accuracy : ±3Bpm
Temperature
Measurement Method : Thermistor Method
Measurement Range : 0 to 50
C / 32 to 122
F
Measurement Accuracy : ± 0.2
C (probe and main unit)
± 0.1
C (main unit only)
SpO
2
(Arterial Oxygen Saturation)
Measurement Method
: 2 Wavelength Pulse Wave Method
Wavelength: Approx. 660nm (red light)
Approx. 890nm (infrared light)
Output: 15mW or below
Measurement Range
: 1 to 100%
Resolution :
1%
Measurement Accuracy : Adult
± 2% when 70 to 100%
(± 3% when DS-100A is used)
Not specified when less than 70%.
Neonate ± 2% when 70 to 100%
Not specified when less than 70%.
PR Measurement Range : 20 to 300bpm
PR Resolution
: 1bpm
PR Accuracy
: ± 3bpm when 20 to 250bpm
Not specified when 251 to 300bpm.
Measurement
Update Rate
: 1 sec.
Measurement
Response Time
: 6 to 7 sec. (averaging duration)
Summary of Contents for DSL-8001
Page 1: ......
Page 2: ......
Page 34: ...xxxii ...
Page 50: ...1 8 ...
Page 56: ...2 6 Setup 2 2 window ...
Page 88: ...2 38 Reference For details refer to 4 Monitoring Setup Printing Setup ...
Page 90: ...2 40 ...
Page 112: ...3 22 ...
Page 166: ...6 10 ...
Page 182: ...7 16 ...
Page 216: ...8 34 ...
Page 250: ...10 6 ...
Page 251: ......
Page 252: ......