The <NIBP Check Cuff, Airhose> message is displayed.
Cause 1 : The connection between the cuff and air hose or the air hose and
NIBP connector is loose or disconnected.
Solution : If connection is loose or disconnected, securely connect it and
perform the measurement again.
If the same message is displayed again, air may internally leak.
Cease the measurement and contact our service representative.
Cause 2 : The cuff is compressed.
Solution : Make sure that the cuff is not subjected to compression, and
measure with the patient arms extended as much as possible.
If the same message is displayed again, a blockage in the air
system can be considered. Cease the measurement and contact our
service representative.
The <NIBP Motion Artifact> message is displayed, and measurement has
stopped.
Cause
: The measurement accuracy is not reliable due to body motion
artifact.
Solution : During the measurement, have the patient stay still as much as
possible.
The <NIBP System Error (E**-**)> message is displayed.
Cause : An error has occurred on the NIBP unit.
Solution : Reboot the equipment. If the same message is repeatedly displayed,
a failure of the equipment can be considered. Cease the
measurement and contact our service representative.
The <NIBP Inflation Timeout (C01)> message is displayed.
Cause 1 : The air hose is not firmly connected, and the air is leaking.
Solution : Check if the air hose is properly connected.
Cause 2 : The cuff size does not match the selected patient type.
Solution : Use the cuff with correct size for the selected patient type.
The <NIBP Cannot Measure (C02)> message is displayed.
Cause 1 : The blood pressure may not be correctly measured due to the
patient’s condition.
Solution : Check the patient’s condition, set the quick measurement to OFF,
and measure again.
Cause 2 : The cuff application may be loose.
Solution : Check that the cuff size is appropriate for the patient, and then
measure again after wrapping the cuff properly.
The <NIBP Exhaust Error (C03)> message is displayed.
Cause 1 : During measurement, an artifact such as body motion may have
interfered.
Summary of Contents for DSL-8001
Page 1: ......
Page 2: ......
Page 34: ...xxxii ...
Page 50: ...1 8 ...
Page 56: ...2 6 Setup 2 2 window ...
Page 88: ...2 38 Reference For details refer to 4 Monitoring Setup Printing Setup ...
Page 90: ...2 40 ...
Page 112: ...3 22 ...
Page 166: ...6 10 ...
Page 182: ...7 16 ...
Page 216: ...8 34 ...
Page 250: ...10 6 ...
Page 251: ......
Page 252: ......