8
-
5
●
Alarm Indicator
Select the alarm level to activate the alarm indicator.
1 Press the
(Menu) key, and select “Maintenance” > enter the
password > “Alarm Behavior Setup” > “Alarm Indicator”.
The “Alarm Indicator”
window will be displayed.
2 Select from [Alarm Level 1-3] / [Alarm Level 1-2] / [Alarm Level 1] / [OFF].
For example, if [Alarm Level 1] is selected, alarm indicator will activate only when
the level 1 alarm occurs.
If [OFF] is selected, alarm indicator will not function.
●
HR Zero Alarm
Select whether or not to generate an alarm when the HR value is 0.
1 Press the
(Menu) key, and select “Maintenance” > enter the
password > “Alarm Behavior Setup” > “HR zero Alarm”.
2 Select
[ON]/[OFF].
[ON] will generate the alarm when the HR value is zero.
[OFF] will not generate the alarm even if the HR value is zero.
NOTE
HR/PR alarm and HR zero alarm is independent and the setting is
not linked with each other. For example, even if HR/PR alarm
setting is OFF, HR zero alarm can be set to ON.