2
-
28
Alarm Function
This section describes about the alarm function.
CAUTION
On a wireless network, the alarm generated on this unit will be
transmitted to the central monitor with a maximum of 15 seconds
delay.
Alarm Level and Alarm Tone
Depending on the alarm level, sound, color, and alarm indicator will differ.
Alarm Level
(Definition)
Sound*
Characters
/Background
Color
Alarm
Indicator
Color
Frequency
of flashing
Level 1
(Life Threatening)
Continuous tone
White/Red
Red 2Hz
Level 2
(Cautionary)
5 seconds
interval tone
Black/
Yellow
Yellow 0.6Hz
Level 3
(Treatment Needed)
Single beep tone
White/Blue
Yellow Lighting
Level 4
(Notification Alarm)
Display Only
Black/
White
- -
* Alarm Sound Pattern (Pattern 1 / Pattern 2) can be selected on the “Sound
Setup” window.
CAUTION
The alarms are classified into Level 1 (High Priority), Level 2
Medium Priority), Level 3 (Low Priority), Level 4 (Notification),
and the alarm message will be displayed according to the
priority of level 1>level 2
>
level 3>level 4.
If more than one alarm is generated, alarm messages will be
displayed by switching the alarm messages in 2-second
intervals.
Summary of Contents for DSL-8001
Page 1: ......
Page 2: ......
Page 34: ...xxxii ...
Page 50: ...1 8 ...
Page 56: ...2 6 Setup 2 2 window ...
Page 88: ...2 38 Reference For details refer to 4 Monitoring Setup Printing Setup ...
Page 90: ...2 40 ...
Page 112: ...3 22 ...
Page 166: ...6 10 ...
Page 182: ...7 16 ...
Page 216: ...8 34 ...
Page 250: ...10 6 ...
Page 251: ......
Page 252: ......