6 Select “Save” to save the channel ID and group ID.
"Password" window will be displayed.
7 Turn the jog dial and enter the password.
8 Press the jog dial to finalize the setup.
Verify the stored channel ID and group ID.
When the setup is complete, the set channel ID will be displayed.
If there is an error with the channel ID, the following message will be displayed.
“HLX Invalid Setting”
: The entered transmitter channel ID exceeds the
allowable range. Set a different ID.
CAUTION
If the temperature unit is “
F” and BP unit is “kPa”, it will be
converted to “
C” and “mmHg” respectively when
transmitted.
On a wireless network, the alarm generated on the bedside
monitor will be transmitted to the central monitor with a
maximum of 15 seconds delay.