2 Select the port for the ventilator (“COM 1” or “COM 2”).
3 Select the connected ventilator (Servo-i/Servo-s).
NOTE
If other function is already assigned to the COM port, it is necessary
to disconnect the communication in order to change the selection.
WARNING
If [OFF] is selected, the ventilator message will not be displayed
and ventilator alarm will not be generated. Make sure the correct
setting is made.
4 Make sure that ventilator alarm is not generated and ventilator status on
the window is
(green).
CAUTION
If the ventilator status is not
(green), the ventilator alarm will
not be notified even when the alarm condition generates.
5 Start the ventilator monitoring.
CAUTION
Check occasionally the communication status of this unit and
the ventilator. Verify that a ventilator alarm is not generated and
that the ventilator display status is
(green).