Electrode Placement
CAUTION
Verify proper electrode placement, lead, waveform size, and
filter mode selection. If not properly selected, it may cause
erroneous detection.
There are 3-electrode, 4-electrode, and 5-electrode applications depending on
the cable type. Also, the displayed lead type can be changed.
For 3-electrode lead
(1 waveform monitoring)
Lead Type I / II / III
Symbol Color
Electrode
Site
R Red
On the right infraclavicular
fossa
L Yellow
On the left infraclavicular
fossa
F Green
On the left midclavicular line,
near the supracrestal line
For 4-electrode lead
(Max. simultaneous 2 waveforms monitoring)
Lead Type I / II / III / aVR / aVL / aVF
Symbol
Color
Electrode Site
R Red
On the right infraclavicular
fossa
L Yellow
On the left infraclavicular
fossa
F Green
On the left midclavicular line,
near the supracrestal line
N Black
On the right midclavicular
line at the same height as F