9
-
3
Classification
Type of protection
against electric shock
: Class I (AC operation)
Internally powered equipment
(battery operation)
Degree of protection
against electric shock
: Type CF Applied Part
Protection against
defibrillation discharge
: Provided
Degree of protection
against ingress of water
: IPX0 (no protection)
Sterilization or
disinfection method
: Cleaning only
Protection against ignition
of flammable gas
: Not provided
Operation Mode
: Continuous operating equipment
Equipment Type for Transfer : Portable Equipment
Power Supply
Voltage
: AC 100 to 240V/DC10.5 to 12.4V
Frequency
: 50Hz or 60Hz
Power Consumption
: 110VA/65W
Usable Life
6 years: According to self-certification.
For usable life of periodic replacement parts, refer to
"8. Maintenance/Periodic Replacement".
Performance
Display
Device
: 7 inch TFT Color LCD
Function Control : Icon
Waveform Trace : Stationary Trace
Sweep Speed
: ECG/SpO
2
(12.5mm/sec, 25mm/sec, 50mm/sec),
Impedance respiration (6.25mm/sec, 12.5mm/sec,
25mm/sec)
Parameter
: ECG, HR, RESP, TEMP, SpO
2
, NIBP, PR
Operation
Jog Dial
: With push switch
Fixed Keys : 5 keys (Alarm Silence, NIBP Start/Stop, NIBP Auto Mode, Home,
Menu)
Print Key
: 1 key
Summary of Contents for DSL-8001
Page 1: ......
Page 2: ......
Page 34: ...xxxii ...
Page 50: ...1 8 ...
Page 56: ...2 6 Setup 2 2 window ...
Page 88: ...2 38 Reference For details refer to 4 Monitoring Setup Printing Setup ...
Page 90: ...2 40 ...
Page 112: ...3 22 ...
Page 166: ...6 10 ...
Page 182: ...7 16 ...
Page 216: ...8 34 ...
Page 250: ...10 6 ...
Page 251: ......
Page 252: ......