HR/PR Setup
This section describes the setup procedure for ECG lead, ECG size, filter mode,
and pacemaker pulse.
ECG Lead
Set the monitoring lead.
1 Press the
(Menu) key, and select “Parameter Setup” > “HR/PR Setup”
> “ECG1 Lead”.
The "ECG1 Lead" window will be
displayed.
2 Select a lead depending on the type of used lead cable.
(I / II / III / aVR / aVL / aVF / V)
3 Set the “ECG2 Lead” using the same procedure.
ECG Size
1 Press the
(Menu) key, and select “Parameter Setup” > “HR/PR Setup”
> “ECG1 Size”.
The "ECG1 Size" window will be
displayed.
2 Select the size of ECG1. (x1/4 / x1/2 / x1 / x2 / x4 / Auto)
When [Auto] is set, the size will be automatically adjusted to optimal size.
3 Set the “ECG2 Size” using the same procedure.