6
-
3
4 Trend Bar
The position and proportion of the stored data are displayed
through the position and width of the slider.
(1)
(3)
(2)
(1) Stored data
(2) Displayed trend data
(3) Unused portion which the data is not stored
5 Next Screen
Displays next trend data.
6 Newest
Screen
Display
Displays the newest trend data.
Setup for Trend Display
The following setup can be performed for the trend display.
Setup Item
Selection
Default
Display Time
Width
1 hour / 2 hours / 4 hours / 8 hours /12 hours /
24 hours
1 hour
Display Scale
HR/PR:
0-100bpm / 0-200bpm / 0-300bpm / Auto
Auto
SpO
2
:
80-100% / 50-100% / 0-100% / Auto
NIBP:
0-100mmHg (13.3kPa) / 0-150mmHg (19.9kPa) /
0-200mmHg (26.6kPa) / 0-300mmHg (39.9kPa)/
Auto
RESP:
0-50Bpm / 0-100Bpm / 0-150Bpm / Auto
TEMP: 30-40
C (86-104
F) / 20-50
C (68-122
F) /
Auto
ON/OFF of
Trend Display
HR/PR, SpO
2
, NIBP, RESP, TEMP
All ON
Data Clear
No/Yes
―
1 Press the
(Menu) key, and select “Data Setup” >“Trend Setup”.
The “Trend Setup” window will
be displayed.