3
-
4
For 5-electrode lead
(Max. simultaneous 2 waveforms monitoring)
Lead Type I / II / III / aVR / aVL / aVF / V
Symbol
Color Electrode
Site
R Red
On the right infraclavicular
fossa
L Yellow
On the left infraclavicular
fossa
F Green
On the left midclavicular line,
near the supracrestal line
N Black
On the right midclavicular
line at the same height as F
C
White
Chest Lead (V1–V6)
Summary of Contents for DSL-8001
Page 1: ......
Page 2: ......
Page 34: ...xxxii ...
Page 50: ...1 8 ...
Page 56: ...2 6 Setup 2 2 window ...
Page 88: ...2 38 Reference For details refer to 4 Monitoring Setup Printing Setup ...
Page 90: ...2 40 ...
Page 112: ...3 22 ...
Page 166: ...6 10 ...
Page 182: ...7 16 ...
Page 216: ...8 34 ...
Page 250: ...10 6 ...
Page 251: ......
Page 252: ......