Maintenance Check
This section explains the daily check and periodic check items of the equipment.
About the Maintenance Check
Periodic inspection must be performed. When reusing the equipment which was
left unused for a while, always check that the equipment operates properly and
safely before use.
Following maintenance check items should be performed for this equipment. To
ensure safety, reliability, and high performance, a “Daily Check” and “Periodic
Check” must be performed. Please be aware that Fukuda Denshi is not liable of
any accidents arising from lack of daily check.
CAUTION
Do not open the main unit housing.
Do not allow alcohol or other liquids to enter the equipment.
●
Daily Check
Perform the following daily check according to the “Daily Check List” on the next
page.
●
Periodic Check
Periodic check of medical electronic equipment is mandatory to prevent failures
and accidents and to ensure safety and reliability.
Periodic check may be performed by each medical institution or by a third party
by concluding a “Maintenance Contract”.
For more details, contact your nearest service representative.
Periodic Replacement
To ensure reliability of safety, function, and performance of this equipment, the
following components must be replaced periodically. When replacing, contact our
service representative.
NIBP Unit
100,000 times of measurement
Battery
1 year or 300 times of discharge and charge
Summary of Contents for DSL-8001
Page 1: ......
Page 2: ......
Page 34: ...xxxii ...
Page 50: ...1 8 ...
Page 56: ...2 6 Setup 2 2 window ...
Page 88: ...2 38 Reference For details refer to 4 Monitoring Setup Printing Setup ...
Page 90: ...2 40 ...
Page 112: ...3 22 ...
Page 166: ...6 10 ...
Page 182: ...7 16 ...
Page 216: ...8 34 ...
Page 250: ...10 6 ...
Page 251: ......
Page 252: ......