Item
Setup Value
Note
Main Menu/Alarm Setup/Alarm Limit Setup
HR/PR Alarm Upper Limit
OFF, 25 to 300 bpm
HR/PR alarm upper limit can be
selected.
HR/PR Alarm Lower Limit
OFF, 20 to 295 bpm
HR/PR alarm lower limit can be
selected.
SpO2 Alarm Upper Limit
OFF, 71 to 100 %
SpO
2
alarm upper limit can be
selected.
SpO2 Alarm Lower Limit
OFF, 70 to 99 %
SpO
2
alarm lower limit can be
selected.
NIBP SYS Alarm Upper
Limit
OFF, 35 to 250 mmHg
(4.6 to 33.3 kPa)
NIBP SYS alarm upper limit can
be selected.
NIBP SYS Alarm Lower
Limit
OFF, 30 to 245 mmHg
(3.9 to 32.6 kPa)
NIBP SYS alarm lower limit can
be selected.
NIBP MAP Alarm Upper
Limit
OFF, 20 to 235 mmHg
(2.6 to 31.3 kPa)
NIBP MAP alarm upper limit can
be selected.
NIBP MAP Alarm Lower
Limit
OFF, 15 to 230 mmHg
(1.9 to 30.6 kPa)
NIBP MAP alarm lower limit can
be selected.
NIBP DIA Alarm Upper
Limit
OFF, 15 to 200 mmHg
(1.9 to 26.6 kPa)
NIBP DIA alarm upper limit can
be selected.
NIBP DIA Alarm Lower
Limit
OFF, 10 to 195 mmHg
(1.3 to 25.9 kPa)
NIBP DIA alarm lower limit can
be selected.
RESP Alarm Upper Limit
OFF, 4 to 150 Bpm
RESP alarm upper limit can be
selected.
RESP Alarm Lower Limit
OFF, 3 to 149 Bpm
RESP alarm lower limit can be
selected.
TEMP Alarm Upper Limit
OFF, 31.0 to 50.0 °C
(87.8 to 122.0 °F)
TEMP alarm upper limit can be
selected.
TEMP Alarm Lower Limit
OFF, 30.0 to 49.5 °C
(86.0 to 121.1 °F)
TEMP alarm lower limit can be
selected.
Main Menu/Parameter Setup/HR/PR Setup
ECG1 Lead
I / II / III / aVR / aVL / aVF / V
ECG1 lead can be selected.
*Unselectable lead will be
displayed in gray.
ECG1 Size
x1/4 / x1/2 / x1 / x2 / x4 / Auto
ECG1 size can be selected.
ECG2 Lead
I / II / III / aVR / aVL / aVF / V
ECG2 can be selected.
*Unselectable lead will be
displayed in gray.
ECG2 Size
x1/4 / x1/2 / x1 / x2 / x4 / Auto
ECG2 size can be selected.
Filter Mode
Diagnosis Mode / Monitor Mode /
ESIS Mode
ECG filter mode can be selected.
HR/PR Source
SpO2 / ECG / Auto
HR/PR source can be selected.
HR/PR Alarm
OFF / ON
ON/OFF of HR/PR alarm
selected.
Pacemaker Pulse
OFF / ON
ON/OFF of pacemaker pulse
display can be selected.
This setting will be automatically
set to “ON” when “ON” is
selected for “Pacemaker” on the
“Patient Setup”.
Alarm Limit Setup
Displays the “Alarm Limit Setup”
window.
Summary of Contents for DSL-8001
Page 1: ......
Page 2: ......
Page 34: ...xxxii ...
Page 50: ...1 8 ...
Page 56: ...2 6 Setup 2 2 window ...
Page 88: ...2 38 Reference For details refer to 4 Monitoring Setup Printing Setup ...
Page 90: ...2 40 ...
Page 112: ...3 22 ...
Page 166: ...6 10 ...
Page 182: ...7 16 ...
Page 216: ...8 34 ...
Page 250: ...10 6 ...
Page 251: ......
Page 252: ......