Patient Neck Count
After the patient background is measured and saved, Figure 6-31 will appear and the
patient’s neck must be counted to determine the activity of isotope concentrated in the
thyroid gland.
Figure 6-31
To measure the patient’s neck, select the
Yes
button. You must make at least one neck
measurement. For multiple repetition protocols, you may omit the second count by selecting
No
. To exit from the measurement process at any time without saving the measurement,
select
Cancel
. After the measurement has been completed, you will be given the
opportunity to reject the count. Figure 6-26 will appear and ask for data verification. Select
YES
to accept the count and save the data to the patient result file, select
NO
to reject the
count data and return to the previous measurement in the counting sequence.
If Decay Correct Administered Dose was selected for the counting protocol, then subsequent
uptake counts on the same patient will automatically calculate the capsule count, and
proceed to Figure 6-30 and 6-31, patient background and neck counts.
If Measure Same Reference Dose Before Each Uptake was selected as the counting
protocol, then each subsequent uptake count of the same patient will require the entire
counting sequence from Figure 6-25 through Figure 6-31 to be repeated. The software does
allow the option to choose to decay correct subsequent counts rather than repeat the
capsule count. To decay correct subsequent capsule counts using Measure Same
Reference Dose protocol, select the
NO
button from the Capsule Count Screen, Figure 6-25.
Figure 6-32 will appear.
Figure 6-32
6-24
THYROID UPTAKE
July 00
Summary of Contents for CAPTUS 2000
Page 2: ......
Page 8: ......
Page 26: ......
Page 47: ...CAPINTEC INC CAPTUS 2000 Figure 4 16 July 00 SYSTEM SETUP 4 15...
Page 48: ...CAPINTEC INC CAPTUS 2000 4 16 SYSTEM SETUP July 00...
Page 52: ......
Page 63: ...CAPINTEC INC CAPTUS 2000 Figure 5 9 July 00 QUALITY ASSURANCE 5 11...
Page 66: ......
Page 105: ...CAPINTEC INC CAPTUS 2000 Figure 7 10 July 00 WIPE TESTS 7 11...
Page 133: ...CAPINTEC INC CAPTUS 2000 Figure 8 31 July 00 LAB TESTS 8 25...
Page 145: ...CAPINTEC INC CAPTUS 2000 Figure 8 46 July 00 LAB TESTS 8 37...
Page 157: ...CAPINTEC INC CAPTUS 2000 July 00 LAB TESTS 8 49 Figure 8 59...
Page 158: ......
Page 165: ...CAPINTEC INC CAPTUS 2000 Figure 9 7 July 00 BIOASSAY 9 7...
Page 173: ...CAPINTEC INC CAPTUS 2000 July 00 BIOASSAY 9 15 Figure 9 17...
Page 174: ......
Page 190: ...CAPINTEC INC CAPTUS 2000 Figure 10 19 10 16 MCA July 00...
Page 198: ...CAPINTEC INC CAPTUS 2000 10 24 MCA July 00 Figure 10 30...
Page 204: ...CAPINTEC INC CAPTUS 2000 11 6 CLEANING AND MAINTENANCE July 00 Figure 11 2...
Page 206: ......