CAPINTEC, INC
CAPTUS
®
2000
Counting Method
First review the counting method box, Figure 6-5. Choose either Decay Correct
Administered Dose or Measure Same Reference Dose Before Each Uptake.
Figure 6-5
When Decay Correct Administered Dose is selected, the capsule must first be measured
before it is administered. Then this same capsule is administered to the patient.
The software will automatically correct the count rate for radioactive decay. Individual
capsules may vary in activity from each other. The decay correct method eliminates any
error introduced by a difference in activity from the capsule administered to the patient, and a
second different capsule used as a standard.
It is the easiest method, since it eliminates the need to recount a capsule each time an
uptake measurement is performed. This method is also the most cost effective since it
eliminates the cost of a second capsule used only as a standard. If your facility uses an
alternate method, try making comparison measurements with your current method and the
decay correct method. Most facilities find that the Decay Correct Administered Dose is the
most convenient method.
When Measure Same Reference Dose Before Each Uptake is chosen, the dosage may be
administered to the patient at any time. Before an uptake measurement is counted, the
software will prompt the user to measure the capsule first, then measure the patient. The
uptake will be calculated from each capsule count. If at any time, you choose not to re-
measure the capsule for subsequent uptake measurements, the software will automatically
decay correct the most recent capsule count for the next uptake calculation.
July 00
THYROID UPTAKE
6-5
Summary of Contents for CAPTUS 2000
Page 2: ......
Page 8: ......
Page 26: ......
Page 47: ...CAPINTEC INC CAPTUS 2000 Figure 4 16 July 00 SYSTEM SETUP 4 15...
Page 48: ...CAPINTEC INC CAPTUS 2000 4 16 SYSTEM SETUP July 00...
Page 52: ......
Page 63: ...CAPINTEC INC CAPTUS 2000 Figure 5 9 July 00 QUALITY ASSURANCE 5 11...
Page 66: ......
Page 105: ...CAPINTEC INC CAPTUS 2000 Figure 7 10 July 00 WIPE TESTS 7 11...
Page 133: ...CAPINTEC INC CAPTUS 2000 Figure 8 31 July 00 LAB TESTS 8 25...
Page 145: ...CAPINTEC INC CAPTUS 2000 Figure 8 46 July 00 LAB TESTS 8 37...
Page 157: ...CAPINTEC INC CAPTUS 2000 July 00 LAB TESTS 8 49 Figure 8 59...
Page 158: ......
Page 165: ...CAPINTEC INC CAPTUS 2000 Figure 9 7 July 00 BIOASSAY 9 7...
Page 173: ...CAPINTEC INC CAPTUS 2000 July 00 BIOASSAY 9 15 Figure 9 17...
Page 174: ......
Page 190: ...CAPINTEC INC CAPTUS 2000 Figure 10 19 10 16 MCA July 00...
Page 198: ...CAPINTEC INC CAPTUS 2000 10 24 MCA July 00 Figure 10 30...
Page 204: ...CAPINTEC INC CAPTUS 2000 11 6 CLEANING AND MAINTENANCE July 00 Figure 11 2...
Page 206: ......