PREVENTATIVE MAINTENANCE
The Quality Assurance Tests described in CHAPTER 5: ACCEPTANCE & QUALITY
ASSURANCE TESTS, should be performed periodically as stated.
Tests must be performed in an environment where the temperature is stable within a range
of +50°F to +85°F (+10°C to +30°C) and the maximum relative humidity is 90% non-
condensing. The unit should be powered-up for at least one-half hour prior to performing any
measurements. No other precautions need to be observed.
CAUTION
If these environmental requirements are not followed, the instrument may display erroneous
readings.
NOTE:
Never put a sharp instrument like a pen into the well or collimator. If the aluminum
coating of the NaI detector is scratched, allowing light and moisture to penetrate, the detector
will be ruined. The rod source is plastic and can be safely placed against the surface of the
detectors when handled carefully.
If the unit fails to pass Autocalibration Test or any other test, the user
should not
attempt to
perform any adjustments to the system. In this event, please contact Capintec for further
assistance.
The Quality Assurance tests should be immediately performed if:
•
the equipment has been subjected to extreme physical stress,
•
liquids enter the readout unit, and/or probe and well detectors, or
•
any cable shows signed of damage
SERVICING
The system is covered by a one year limited warranty, under normal conditions of use.
Other than the printer’s ink cartridges and the isolation transformer’s fuses (see Fuse
Servicing in this chapter), there are no user serviceable parts contained in the system.
Every five years, the system should be returned to the manufacturer for a complete
verification.
DISPOSAL
The following items should be taken into consideration before disposing.
The unit contains the following types of hazardous items:
•
Lead – the shielding around the probe and well detectors and inside of the Spring-
Arm stand.
•
Lithium Battery – contained in the Computer.
These items should be disposed of in accordance with local and national regulations.
July 00
CLEANING AND MAINTENANCE
11-3
Summary of Contents for CAPTUS 2000
Page 2: ......
Page 8: ......
Page 26: ......
Page 47: ...CAPINTEC INC CAPTUS 2000 Figure 4 16 July 00 SYSTEM SETUP 4 15...
Page 48: ...CAPINTEC INC CAPTUS 2000 4 16 SYSTEM SETUP July 00...
Page 52: ......
Page 63: ...CAPINTEC INC CAPTUS 2000 Figure 5 9 July 00 QUALITY ASSURANCE 5 11...
Page 66: ......
Page 105: ...CAPINTEC INC CAPTUS 2000 Figure 7 10 July 00 WIPE TESTS 7 11...
Page 133: ...CAPINTEC INC CAPTUS 2000 Figure 8 31 July 00 LAB TESTS 8 25...
Page 145: ...CAPINTEC INC CAPTUS 2000 Figure 8 46 July 00 LAB TESTS 8 37...
Page 157: ...CAPINTEC INC CAPTUS 2000 July 00 LAB TESTS 8 49 Figure 8 59...
Page 158: ......
Page 165: ...CAPINTEC INC CAPTUS 2000 Figure 9 7 July 00 BIOASSAY 9 7...
Page 173: ...CAPINTEC INC CAPTUS 2000 July 00 BIOASSAY 9 15 Figure 9 17...
Page 174: ......
Page 190: ...CAPINTEC INC CAPTUS 2000 Figure 10 19 10 16 MCA July 00...
Page 198: ...CAPINTEC INC CAPTUS 2000 10 24 MCA July 00 Figure 10 30...
Page 204: ...CAPINTEC INC CAPTUS 2000 11 6 CLEANING AND MAINTENANCE July 00 Figure 11 2...
Page 206: ......