Figure 8-37
Note the
Go To MCA
command located at the top of the Blood Volume Data Acquisition
screen. This features provides a direct link to the complete MCA program without exiting
from the Blood Volume module. Use this feature to check the isotope identification of a
source, verify a region of interest, check a patient for previously administered isotopes, or
any other MCA use.
A sequence screens will appear and prompt you through the correct counting sequence. As
the measurement process begins and counts are acquired, the live spectra will be displayed
on the screen. The ROI selected will be highlighted in red to easily verify that it is correct.
The live time remaining will appear in the upper right hand corner of the screen, while peak
counts will appear in the left hand corner of the screen.
To abort the count, select the
ABORT
ACQUISITION
button, which appears during live
acquisition, and the count will not be saved. To stop the count and save the counting data
before the preset live time has elapsed, choose
STOP & SAVE
. When you are finished
counting, select the
EXIT
command. The display will return to the Main Blood Volume Test
screen. If two repetitions have been selected in the counting protocol, each of the
subsequent count screens will appear twice and two measurements for each count will be
recorded. The results of the two counts will be averaged. Both the raw count data and the
July 00
LAB TESTS
8-31
Summary of Contents for CAPTUS 2000
Page 2: ......
Page 8: ......
Page 26: ......
Page 47: ...CAPINTEC INC CAPTUS 2000 Figure 4 16 July 00 SYSTEM SETUP 4 15...
Page 48: ...CAPINTEC INC CAPTUS 2000 4 16 SYSTEM SETUP July 00...
Page 52: ......
Page 63: ...CAPINTEC INC CAPTUS 2000 Figure 5 9 July 00 QUALITY ASSURANCE 5 11...
Page 66: ......
Page 105: ...CAPINTEC INC CAPTUS 2000 Figure 7 10 July 00 WIPE TESTS 7 11...
Page 133: ...CAPINTEC INC CAPTUS 2000 Figure 8 31 July 00 LAB TESTS 8 25...
Page 145: ...CAPINTEC INC CAPTUS 2000 Figure 8 46 July 00 LAB TESTS 8 37...
Page 157: ...CAPINTEC INC CAPTUS 2000 July 00 LAB TESTS 8 49 Figure 8 59...
Page 158: ......
Page 165: ...CAPINTEC INC CAPTUS 2000 Figure 9 7 July 00 BIOASSAY 9 7...
Page 173: ...CAPINTEC INC CAPTUS 2000 July 00 BIOASSAY 9 15 Figure 9 17...
Page 174: ......
Page 190: ...CAPINTEC INC CAPTUS 2000 Figure 10 19 10 16 MCA July 00...
Page 198: ...CAPINTEC INC CAPTUS 2000 10 24 MCA July 00 Figure 10 30...
Page 204: ...CAPINTEC INC CAPTUS 2000 11 6 CLEANING AND MAINTENANCE July 00 Figure 11 2...
Page 206: ......